Scroll through the results on the main display with the R-DAR-PI-DD button, or on
the small display with the V/TIME button.
Use R(t) when carrying out a 'time-controlled test' to scroll through the interval
measurement values saved at the sampling rate specified in SET-UP. These
values are available until another measurement is taken or the selector switch is
turned.
When measuring high levels of insulation (> 1GΩ), it is advised that the "G"
guard terminal be used to eliminate the influence of surface leakage currents. The
guard is connected between the two measurement contact points, and the surface
susceptible to surface currents, (e.g. dusty, damp cable or transformer insulation).
In this case, alligator clips are preferable to test-probes held in the hand.
As soon as insulation measurement is stopped, the test circuit is automatically
discharged using the instrument's internal resistance.

5.3 Operation Examples

Connection diagram for measurement of low insulation (e.g. motor)
Connection diagram for measurement of strong insulation
a) Example of a motor (reduction of capacitive effects)
34
0.561.8187
M
M
www.
+
110 - 230V
50/60 Hz
20 VA max
SET-UP
MΩ 50-5000V
MΩ - 5000V
MΩ - 2500V
FF 0.1A
380V - 10kA
5x20mm
MΩ -1000V
MΩ - 500V
G
OFF
V/TIME
-
SMOOTH
ALARM
MEM
PRINT
R-DAR-PI-DD
1000V CAT III
(
)
2500V
2
nd
MR
PRINT MEM
R(t)
+
110 - 230V
50/60 Hz
20 VA max
G
FF 0.1A
380V - 10kA
5x20mm
G
-
SMOOTH
ALARM
MEM
PRINT
1000V CAT III
(
2500V
)
2
nd
MR
PRINT MEM
Megohmmeter Models 5050/5060
.com
RS-232
START/STOP
MEGOHMMETER
MODEL 5060
RS-232
SET-UP
MΩ 50-5000V
MΩ - 5000V
MΩ - 2500V
MΩ -1000V
MΩ - 500V
OFF
V/TIME
START/STOP
R-DAR-PI-DD
MEGOHMMETER
R(t)
MODEL 5060
information@itm.

Advertisement

Table of Contents
loading

This manual is also suitable for:

5060

Table of Contents