Appendix C-Ieee-488 - Keithley 199 Instruction Manual

System dmm scanner
Hide thumbs Also See for 199:
Table of Contents

Advertisement

APPENDIX C
Figure
C-2. IEEE Handshake
Sequence
BUS COMMANDS
Commands
associated
with the IEEE-488
bus can be
grouped into the following three general categories. Refer
to Table C-l.
IJniIine CommandsThese
commands are asserted by set-
ting the associated
bus line true. For example, to assert
REN (Remote Enable),
the REN line would be set low
&W.
MultiIine
Commands-General
bus commands which are
sent over the data lines with the ATN line true.
Device-dependent
Commands-Commtids
whose mean-
ings depend on the device in question.
These commands
are transmitted
via the data lines while ATN is false.
Table C-l. IEEE-488
Bus Command
Summary
State of
Command
Type
Command
ATN Lie*
Comments
Uniline
REN (Remote Enable)
X
Sets up devices for remote operation.
EOI
X
Marks end of transmission.
IFC (Interfack Clear)
X
Clears Interface.
AIN (Attention)
Low
Defines
data bus contents.
SRQ
X
Controlled
by external device.
Multiline
Universal
iL0
(Local Lockout)
Low
Locks out local operation.
DCL (Device Clear)
Low
Returns device to default conditions.
SPE (Serial Enable)
Low
Enables serial polIing.
SPD (Serial PoU Disable)
Low
Disables
serial polling.
Addressed
SDC (Selective
Device Clear)
Low
Returns unit to default conditions.
GTL (Go To Local)
IDN
Returns device to local.
GET (Group Execute Trigger)
Low
Triggers device for reading.
Unaddressed
UNL (Unlisten)
Low
Removes all listeners from bus.
UNT (Untalk)
Low
Removes any talkers from bus.
Device-dependent
High
Programs Model 199 for various modes.
*Don't Care.
c-3

Hide quick links:

Advertisement

Table of Contents
loading