SECTION
4
Performance Verification
4.1 INTRODUCTION
4.3 INITIAL CONDITIONS
The procedures
outlined
in this section may be used to
verify that the instrument
is operating
within the limits
stated in the specifications
at the front of this manual. Per-
fotiance
verification
may be performed
when the instru-
ment is first received to ensure
that
no damage or misad-
justment has occurred during shipment.
Verificatioiimay,
also be performed
whenever
there is a question of instru-
ment accuracy, or following
calibration,
if desired.
The Model 199 must be turned on and allowed to warm
up for at least two hours before beginning the verfication
procedures.
If the instrument has been subject to extremes
of temperature
(outside the range specified in paragraph
4.2),
additional
time
should
be allowed
for internal
temperatures
to reach
normal
operating
temperature.
Typically, it takes one additional
hour to stabilize a unit
that is 10°C (l8"F) outside the specified temperature range.
NOTE
If the instrument
is still under warranty (less than
4.4 RECOMMENDED
TEST EQUIPMENT
1 year from the date of shipment),
and its perform-
ance falls outside the specified rage,
contact your
Table 4-l lists all test equipment
required for verification.
Keithley representative
or
the
factory to determine
Alternate equipment
may be used as long as the substitute
the correct course of action.
equipment has specifications
at least as good as those listed
ifi the table.
4.2 ENVIRONMENTAL
CONDITIONS
All measurements
should be made at 18 - 28°C (65 - 82°F)
and at less than 80% relative humidity.
NOTE
The verification limits in this section do not include
test equipment
tolerance
and are based on one
year accuracy specifications.
Table 4-1. Recommended
Test Equipment
Mfg
1
Model
Description
Specifications
I
k
I
Fluke
Fluke
Fluke
Fluke
Valhalla
544OA
DC Voltage Calibrator
3OOmV, 3v, 3ov, 3oov ranges +15ppm.
5200A
AC Voltage Calibrator
3OOmV, 3V, 30V ranges; 2OH.z &O.l%;
50Hz-20kHz
0.02%; lOOkI+ io.33%.
5215A
AC Power Amplifier
3COV range: 2OHz &X2%;
SOHz-20kHz 10.04%;
lOOkHzfo.l%
5450A
Resistance Calibrate?
3ooR-3MQ
ranges i-15ppm; 3OMQ L32ppm;
300MR range X?2Sppm
2500E
AC-IX
Ciarrent Calibrator
3OmA, 3A ranges M.03% DC, io.l%
AC to 5kHz
4-l