199 IEEE-488 Programming
DEVICE-DEPENDENT
COMMANDS
EXECUTE
X
Execute other device-dependent
commands
FUNCTION
Fo
DC volts
R
AC volts
F2
Ohms
53
DC current
E
AC current
ACV dB
F6
ACA dB
RANGE
A'3 ACA
DCV
ACV
DCA
ACA
Ohms
dB
dB
Ro
Auto
Auto
Auto
Auto
Auto Auto Auto
s
3OOmV 3WmV 3OmA 3OmA 300 d
'Auto A&
3V
3V
3A
3A
3kl-i Auto Auto
R3
306'
3OV
3A
3.4
3Okn Auto Auto
R4
300V
300V
3A
3.4
3OOM
Auto
Auto
z
3oOV
30%'
3A
3.4
3Mn
Auto
Auto
3mv
3wv
3A
3A
30MR
Auto
Auto
R7
3OOV
3wV
3A
3A 3OOMO
Auto
Auto
ZERO
Zero disabled
Zero enabled
Zero enabled using a zero value
(V)
FILTER
PO
Pl
P2
Internal and front panel filter disabled
Internal filter enabled
Front panel filter enabled
RATE
so
Sl
4%digit
resolution,
2.59msec integration
period
5%digit
resolution,
lie
cycle integration
(16.67msec. 6OHz; 2Omsec, 50Hz)
TRIGGER
MODE
m
Continuous on Talk
n
One-shot
on Talk
T2
Continuous
on GET
T3
One-shot
on GE?
T4
co"tin"o"s
on x
T5
T6
T7
One-shot
on X
Continuous
on External Trigger
One-shot
on External
Trigger
READING
BO
z
MODE
Readings from
AID converter
Individual
readings from data store
All readings from data stwe (buffer dump)
DATA STORE
SIZE
IO
Wrap around data store mode
h
Data store of n (n=I to 500)
INTERVAL
Q'J
Qn
Default interval,
l75msec
(SELECT OFF)
n=intwval
in milliseconds
(l5msec to
999999msec)
VALUE
v*nn.nnnn or
Calibration
value, zzo value
v*n.nnnnnrtE+n
CALIBRATION
co
Calibrate first
point using value (V)
2
Calibrate second point using r&e
(V)
Cd&rate
third point using value (V)
DEFAULT CONDITIONS
Lo
Restore factory default conditions
and save (Ll)
L*
Saw present machine
states as default conditions
DATA FORMAT
GO
Reading with prefix.
Gl
Reading without prefu.
G2
Reading and buffer location with prefix.
G3
Reading and buffer I&ion
without prefix.
G4
G5
Reading and channel
with prefix.
Reading and channel without
prefix.
G6
G7
Reading, buffer location,
and channel with prefix
Reading, buffer location, and channel without
prefix
SRQ
MO
Disable
Ml
Reading overflow
M2
Data stcre full
M4
Data store half full
MB
Reading done
Ml6
Ready
M32
Fmx
EOI AND BUS HOLD-OFF
Ko
Enable EOI and bus hold-off on X
E
Disable SO,, enable bus hold-off on X
Enable EOI, disable bus hold-off on X
K3
Disable both EOI and bus hold-off on X
TERMINATOR
MI
ii
Y2
n
CR LF
LFCR
STATUS
UO
Ul
uz
u3
u4
u5
MULTIPLEX
A0
Al
DELAY
Wll
SELFTEST
JO
Send machine status word
Send error conditions
Send Translator
word list
Send buffer size
Send current value of '?I'
Send input switch status (front/rear)
AukKzd
multiplez
disabled
Auto/Cal multiplex
enabled
n=delay period in milliseconds,
(Omsec
to
999999msec)
Test,
ROM,
RAM, E?'ROM
HIT BUTTON