Basicmeas-Ijrements; Warmupperiod; Zero - Keithley 199 Instruction Manual

System dmm scanner
Hide thumbs Also See for 199:
Table of Contents

Advertisement

BASIC DMM OPERATION
2.6 BASIC MEASUREMENTS
The following paragraphs
describe the basic procedures
for making voltage, resistance,
current, and dB measure-
ments.
High Energy Circuit Salty
Precauticms
To optimize safety when measuring voltage in high energy
distribution
circuits, read and use the directions
in the
following warning.
WARNING
Dangerous arcs of an explosive nature in a high
energy circuit can cause severe personal injury
or death. If the meter is connected
to a high
energy circuit when set to a current range, low
resistance
range or any other low impedance
range, the circuit is virtually shorted. Dangerous
arcing can also result when the meter is set to
a voltage range if the minimum
voltage spac-
ing is reduced.
When making measurements
in high energy circuits use
test leads that-meet
the following requirements:
l
Test leads should be fully insulated.
l
Only use test leads that can be connected
to the circuit
(e.g. alligator
clips,
spade
lugs,
etc.)
for hands-off
measurements.
l
Do not use test leads that decrease voltage spacing. This
diminishes
arc protection
and
creates
a hazardous
condition.
Use the following sequence
when testing power circuits:
1.
2.
3.
4.
5.
6.
De-energize
the circuit
using
the regular
installed
connect-disconnect
device such as the circuit breaker,
main switch, etc.
Attach the test leads to the circuit under test. Use ap-
propriate safety rated leads for thii application.
Set the DMh4 to the proper function
and range.
Energize
the
circuit
using
the
installed
connect-
disconnect
device and make measurements
without
disconnecting
the DMM.
De-energize
the circuit using the installed
connect-
disconnect
device.
Dkonnect
the test leads from the circuit under test.
CAUTION
CAUTION
The maximum
common-mode
input voltage
The maximum
common-mode
input voltage
(the voltage
between
input LO and chassis
(the voltage
between
input LO and chassis
ground) is 500V peak. Exceeding this value may
ground) is 500V peak. Exceeding this value may
damage the instrument.
damage the instrument.
2.6.1 Warm Up Period
The Model 199 is usablq~jmmediately when it is f&t turned
on. Howeve!, the instrument must be aowed
to warm up
for at least two hours to achieve rated accuracy.
2.6.2 Zero
The zero feature serves as a means of baseline %uppres-
sion by allowing a stored offset value to be subtracted from
subsequent
readings. When the ZERO button is pressed,
the instrument
takes the currently
displayed reading as
a baseline value. All subsequent readings represent the dif-
ference between
the applied signal level and the stored
baseline.
A baseline level can be established
for any or all measure-
ment functions
and is remembered
by each function.
For
Ample,
a 1OV baseline can be established
on DCV, a 5V
baseline can be established
on ACV and a 1OkQ baseline
can be established on OHMS at the same time. These levels
will not be cancelled by switching back and forth between
functions.
Once a baseline
is established
for a measure-
ment function, that stored level wiU be the same regardless
of what range the Model 199 is on. For example, if 1V is
established
as the baseline
on the 3V range,
then the
baseline will also be 1V on the 30V through 3oOV ranges.
A zero baseline
level can be as large as full range.
NOTE
The following
discussion on dynamic
range is
based on a disolav resolution of 5% dieits. At 41/zd
"
resolution,
th'e Lumber
of counts
would be re-
duced by a factor of 10.
By design, the dynamic measurement
range of the Model
199, at 5%digit
resolution,
is 606,OKl counts. With zero
disabled,
the displayed reading range of the instrument
is GO3,OCHl counts. With zero enabled,
the Model 199 has
the capability to display ~606,OCKl c&n-&. This increased
display range
ensures
that~~the dynamic
measurement
range of the instrument
is not reduced when using a zero
2-7

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents