Resistance Measurements; Two-Terminal Resistance Measurements; Four-Terminal Resistance Measurements; Resistance Ranges - Keithley 199 Instruction Manual

System dmm scanner
Hide thumbs Also See for 199:
Table of Contents

Advertisement

BASIC DMM OPERATION
2.6.6 Resistance Measurements
The Model 199 can make resistance
measurements
from
lmf? to 300MQ. The Model 199 provides automatic selec-
tion of Z-terminal or 4terminal
resistance
measurements.
Th% means that if the ohms sense leads are not connected,
the measurement
is done 2-terminal. If the sense leads are
connected,
the measurement
is done 4terminal.
For
4-terminal measurements,
rated accuracy can be obtained
as long as the maximum lead resistance
does not exceed
the values listed in Table 2-3. For best results on the 3WQ.
3k0 and 3OkQ ranges, it is recommended
that 4terminal
measurements
be made to eliminate errors caused by the
voltage drop across the test leads which will occur when
MODEL
199
----
Figure 2-5. Four-Terminal Resistance
Measurements
2-ten&al
meazuements
are made.
The Model
5806
Kelvin Test Lead Set is ideal for low resistance 4termina.l
measurements.
In the 4% digit mode, use 4terminsl
or
Notes:
connect the source leads to the sense leads at the instru-
ment to avoid extra noise pickup.
1. Table 2-3 shows the current output for each resistance
range.
To ~make resistance
measurements,
proceed as follows:
2. It helps
to shield
resistance
greater
than
1OOkQ (or
anytime 4% digit resolution
is used) to achieve a stable
1. Select the ohms function by pressing the QHMS buttqn.
2. Select a range consistent
with the expected resistance
or use autoranee.
"
3. Select the front or rear panel input termirials using the
J.NruT switch.
4. For 2-terminal
measurements
connect the resistance
to
the instrument
as shown in Figure 2-4. For 4terminal
measurements
connect the resistance
to the instrument
as shown in Figure 2-5.
5. Take the reading from the display.
reading. Place the resistance in a shielded enclosure and
electrically
connect the shield to the LO input terminal
of the instrument.
3. Diode Test-The
3kQ range can be used to test diodes
as follows:
A. Select the 3kR range.
8. Forward bias the diode by connecting
the red ter-
minal of the Model 199 to the diode anode. A good
diode will typically measure between 3GGQ and IkO.
C. Reverse bias the diode by reversing the c+u-wtions
on the diode.
A good diode
will overrange
the
display.
Shielded
Optional shield
Table 2-3. Resistance Ranges
Maximum
Test Lead
S=hd
Nominal
Resistance
(Q) for
Range
Resolution
I-Short
~1 Count Error (Wzd)
300
0
hnfl
1.7mA
10
3kfl
lOmQ
1.7mA
30
30kR
lOOmdt
le.4
100
Figure 2-4. Two-Terminal Resistance Measurements
300kR
10
5w
300
3MQ
10
n
$A
Ik
30MQ
100
n
O&A
3k
3ooMO
1 kO
0.5&4
30k
NOTE: Typical open circuit voltage is 5.511
2-11

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents