Ohms Verification; Connections For Trms Ac Volts Verification; Limits For Trms Ac Volts Verification - Keithley 199 Instruction Manual

System dmm scanner
Hide thumbs Also See for 199:
Table of Contents

Advertisement

Table 4-3. Limits for TRMS AC Volts Verification
199
Applied
Allowable Readings
WC
to '28Tl
ACV Range
AC Voltage
ZOHZ
5oHz
200Hz
lOkHz
1OOkHz
3OOmv
29O.OOOmv
284.100
288.885
289.365
289.365
283.900
29&l
29k5
Z&5
29&5
29&W
3 v
2.90000V
2.84lOO
2.88885
2.89365
2.89365
2.85350
2.9&Q
Z.%l5
2.600635 2.&35
2.94650
to
30
v
29.MxMv
28.4100
28.8885
28.9365
28.9365
28.5330
29:090
29%5
29.:35
29.k.5
29.4650
to
3ooV
29O.OOOV
284.100
288.885
289.365
289.365
*
to
to
to
to
295.900
291.lS
290.635
290.635
*Do not apply 290V at 1OOkHz to the input. This exceeds the V*&
limit of the
instrument.
Maximum TRh4S AC volt input at 1CHlkHz is 1WV. On the SHIV range,
allowable readings with lOOV @ lOOkI& applied to the input are 98.200 to lO'L800.
See par+aph
2.6.7 for clarification
of the V-Hz
specification.
Figure 4-2. Connections for TRMS AC Volts
Verification
4.5.3 Ohms Verification
With the Model 199 set to 5%d resolution,
verify the ohms
function as follows:
CAUTION
Do
not exceed 425V peak or 300V RMS between
the input HI and LO terminals or damage to the
instrument
may occur.
1. Select the ohms function
and autorange.
2. Using Kelvin test leads (such as the Keithley Model 1641)
connect the resistance
calibrator
to the Model 199 as
shown in Figure 43.
3. Set the calibrator to the SHORT position and enable zero
on the Model 199. Verify that the display reads ooO.000
4. Set the calibrator
to output 1900 and verify that~ the
reading is within the limits listed in Table 4-4.
5. Disable zero and leave it disabled for the remainder
of
the ohms verification
procedure.
6. Utilizing
Figures 43 "d
44,
check the 3kX7 through
3CilMQ ranges by applying
the respective
resistance
levels listed in Table 44.
Verify~ that the readings
are
within the limits listed in the table.
,,
4-3

Hide quick links:

Advertisement

Table of Contents
loading