MAINTENANCE
1.~ Select the ohms function
and the 3COQ range.
Ed. Repeat
steps
A through D for the remaining
ohms
2. Connti
the resistanw
calibrator to the instkment
as
shown in Figure 6-3.
3.20~ front panel calibration,
press SHIFT'LCKAL
and
proceed as follows:
A. With the lYO.OOOQ calibration
displayed
on the
Model 199, set the resistance
calibrator to 190a.
8.
After allowing
sufficient
time-foi
fhe~ calibrator
resistance
to settle, press the NEXT button.
The
following
message
will be displayed
for several
seconds:
WORKING
C. With the OOOlMOQ calibration
point displayed,
set
the resistawe
calibrator
to SHORT
(On).
D. After waiting
sufficient
time
for the calibrator
resistance
to settle. mess the ENTER button.
The
following
message
'will Abe displayed
fro several
seconds:
raiges
w&g
Table 6-g as a guide.
I
Oumut HI
n,
4
Resistance
Shielded
Cable
Figure 6-3. Four-Wire Resistance
Calibration
Configuration
(300S30kQ
Ranges)
ohms rang& using Table 6-6 as a guide.
4. For IEEE-488 bus calibration,
proceed as follows:
A. Set the resistance
calibrator to lYO'2.
B. After allowing
sufficient
time for the resistance
calibrator to settle, send the following
commands
over the
bus: VlYOXCOX.
WORKING
E.
The instrument wi!.l tit
the calibration program and
iim-
Output HI
return to the 300R range.
E
Repeat the procedures
in step 3 for the remaining
r----
2
-1
Resistance
,
Calibrator
,
Model 5450A
1
L-----J
C. Set the resistance
calibrator to SHORT
(On).
I
D. After allowing
sufficient
time for the resistance
calibrator to settle, send the following command_over
th-e bus:VOXCK
Both calibration
constants
will be
Figure 6-4. Two-Wire Resistance
Calibration
automatically
stored in E'I'ROM.
Configuation
(300kQ-300MQ
Ranges)
6-7