Dc Current Verification - Keithley 199 Instruction Manual

System dmm scanner
Hide thumbs Also See for 199:
Table of Contents

Advertisement

PERFORMANCE
VERIFICATION
Table 4-4. Limits for Ohms Verification
set up
Figure 4-3
Figure 4-3
Figure 43
Figure 4-4
Figure 4-4
Figure 4-4
Figure 4-4
190.000 kO
189.948 to 190.052
1.9OOOOMQ
1.89940 to 1.9oo60
19.0OGlMQ
'lg.9767 to 19.0233
lCO.CGOMIt
97.995 to 102.005
4.54 DC Current Verification
With the instrument
set to 5%d resolution,
verify the DC
Resistance
current function as follows:
Sense PHI
Calibrator
sense LO
I
Model S45OA
-
ou@"tLO
'
MODEL ISO
CAUTION
Do not exceed 3A to the AMPS and LO input ter-
minals or the front panel current fuse will blow.
1
" .,,, .,
.,
Figure 4-3. Connections
for Ohms Verification
1. Select the DC4 function
and autoranze.
2. Connect the DC~ctieW
calibration s&ce
to the Model
.
-
199asshownmF~gure43.
3. Set the calibration
source to output +3OmA and verify
that the reading is within the limits listed in Table 4-5.
4. Repeat the procedure for the 3A range by applying the
DC current level listed in Table 45.
Check to see that
the reading is within the limits listed in the table.
5.~ Repeat the procedure
for each of the two ranges with
negative current levels.
Table 4-5. Limits for DC Current Verification
Applied
Allowable
Readings
DC CUrrent
OPC to 28'0
3om.4
3cLoooomA
29.9835 to 30.0165
I
,I
3
A
1 3.00000
A
/ 2.99685
to 3.00315
I
Figure 4-4. Connections
for Ohms Verification
(300kG-3OOMil
Ranges)
4-4