BASIC DMM OPERATION
2.7.6 LED lest
This program allows you to test all the front panel annun-
ciators and LED display segments
to check for proper
operation.
Proceed as follows:
1. Press DMM S!CTUP and then NEXT repeatedly until the
following message is displayed:
LEDS OFF
2. To test the LEDs, use uprange or downrange
to toggle
the display to the following:
LEDS ON
3. Press NEXT to initiate the test.
4. During the test, the instrument
will turn on all the an-
nunciators
and
walk
through
the
various
display
segments
and complete
displays
to verify that ally are.
operating
properly. Following the test, the instrument
will scroll to the next program.
2.7.7 Debug
The debug program is intended to switch various LED%
relays, and logic levels to allow signal tracing through the
instrument duing troubleshooting.
Also, memory tests are
performed.
For complete details on using the debug, refer
to paragraph
6.7.3.
To exit the diagnostic
program,
press any key except
TRIGGER.
2.7.8 Reset
The reset program restores iristrument
setup parameters
to the factory default conditions listed in 'Tables 2-l and 3-7
Perform the following steps to use this program.
1. Press SHIFT DMM SETUP and then NEXT repeatedly
until the following message is displayed:
RESET NO
3. Press NEXT to reset the instrument,
which will return
to the default conditions
listed in Table 2-1.
NOTES~:
1. The reset program can be aborted by pressing
NEXT
with the 'XESET
NO" message
displayed.
2. Once the instrument
has been reset to default condi-
tions, use save setup to save that configuration
if you
desire that the instrument power up in those conditions.
3. The reset program has no effect on the programmed
IEE~l%488 primary address or line frequency
setting.
2.8 FRONT PANEL TRIGGERING
The following paragraphs
discuss front panel triggering,
trigger mode selection, as well as trigger delay acd reading
interval programming.
2.8.1 Trigger Mode Selection
The Model 199 may be operated in two basic trigger modes:
one-shot and continuous. In the one-shot mode, a separate
trigger is required to initiate each reading.
For the con-
tinuous mode, however, only a single trigger is required,
with the conversion rate determined
by the programmed
reading interval. The continuous
trigger mode is the fac-
tory default.
To check or change the selected trigger mode, proceed as
follows.
1. Press SHIFT TRIG SETUP. The instrument
will display
the presently selected trigger mode. For the continuous
mode, the display will show:
coNTINuous
2. For the one-shot
mode, the display
reads:
O~NE SHOT
3. TCtij@e
the trigger mode, press uprange or dowruange.
4. Once the desired trigger mode is displayed, press NEXT
to scroll to the next menu selection
(trigger delay), or
press NEXT to relurn to normal fronts panel display.
2.&e
uprange
or downrange
to toggle the display as
follows:
RESFT YES
2-19