Special Handling; Troubleshooting; Recommended Test Equipment - Keithley 199 Instruction Manual

System dmm scanner
Hide thumbs Also See for 199:
Table of Contents

Advertisement

MAINTENANCE
6.6 SPECIAL HANDLING
OF STATIC-
SENSITIVE
DEVICES
CMOS devices operate at very high impedance
levels for
low power consumption.
As a result, any static charge that
builds up on your person or clothing may be sufficient to
destroy these devices,
if they are not hanged
properly.
CAUTION
In order lo avoid possible
damage, assume that
all devices are static sensitive.
When handling these devices, use the following precau-
tions to avoid damaging
them.
1. The devices listed in the replaceable parts list should be
transported
and handled
only in containers
specially
designed to prevent static build-up. Typically, these parts
will be received
in anti-static
containers
of plastic or
foam. Keep these devices in their original containers un-
til ready for installation.
2. Remove the devices from their protective cbnt%iers
only
at a properly
grounded
work station.
Also ground
yourself with a suitable wrist strap.
3. Handle the devices only by the body; do~not touch the
pins.
4. Any printed circuit board into which the devices is to
be inserted must also be grounded to the bench or table.
5. Use only anti-static
type solder suckers.
6. Use only grounded
tip solder irons.
7. Once the device is installed on the PC board, it is nor-
mally adequately
protected and normal handling
can
resume.
6.7 TROUBLESHOOTING
The troubleshooting
information~contained
in this section
is intended for use by qualified personnel having a basic
understanding
of analog and digital circuitry.
The in-
dividual should also be experienced
at using typical test
equipment as well as ordinary troubleshooting
procedures.
The information
presented here has been written to assist
in isolating a defective circuit or circuit s&ion.
Tsolation
of the specific component
is left to the technician.
Note
that schematic
diagrams and component
location
draw-
ings, which are an essential
aid in troubleshooting,
are
located at the end of Section 7.
6.7.1 Recommended
Test Equipment
Success
in troubleshooting
complex
equipment
like the
Model 199 depends not only on the skill of the technician,
but relies on the use of accurate, reliable test equipment.
Table 6-10 lists the equipment
recommended
for trouble-
shooting the Model 199. Other equipment
such as logic
analyzers
and capacitance
meters could also be helpful,
especially
in difficult situations.
6.7.2 Power Up Self Test
Upon power up, the instrument
will do a number of~tests
on it&.
Testsare performed on memory (ROM, RAM and
EV'ROM).
Whenever
one of the memory tests fails, the
instrument
wiIl lock up and musty be repaired.
If all the
tests pass,
then the frequency
will be displayed.
Im-
mediately after turning on the Model 199, the following
sequence
will take place:
1. A digital check will be performed on the ROM circuitry.
If the test fails, the instrument
will lock up with the
following message
displayed:
0000000000
Replacing
U4 tiay correct the problem.
2. If the ROM test passes,
a digital self-test will then be
performed
on the RAM circuitry. If the RAM test fails,
the instrument
will lock up with the following
error
message
displayed:
AAAAA&%AAA
Replacing
U9 may solve the problem.
3. Finally, the E'PROM circuitry where the calibration con-
stants, IEEE address
and lie
frequency
settings
are
stored will be checked.
If thii test fails, the following
message
displayed:
UNCAL
Pressing any front panel momentary
button will return
the instrument to normal operation however, acomplete
calibration will have to be performed,
and the IEEE ad-
dress and line frequency
setting should be checked.
If
the test still fails after calibration,
try replacing
U20
before recalibrating
the instrument
again.
~4. The instrument
will then go to the power up default
function.
6-17

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents