3.8 GENERAL BUS COMMAND
PROGRAMMING
Generd bus commands are those commands such as DCL
that have the same general purpose regardless of the in-
strument.
Con%naxids supported
by the Model 199 are
summarized
in I3ble 3-6, which lists BASIC 4.0 statements
necessary
to send each command.
Note that commands
requiring
a primary address
assume that the Model 199
primary address is set to 26 (its factory default~address).
3.8.1 REN (Remote Enable)
REN is a uniline command
that must be asserted by the
controller
to place the Model 199 in remote. Simply set-
ting EN
true will not adually place the instrument
in
remote; instead, the unit must be addressed to listen after
REN is set true.
Generally,
remote enable
should be asserted
before at-
tempting to program the instrument
over the bus. Once
the instrument
is in remote, all front panel controls except
LOCAL~will be inoperative.
Normal front panel operation
can be restored by pressing
the LOCAL key.
To place the Model 199 in remote, the controller must per-
form the following
sequence:
1. Set the REN line true.
2. Address
the Model 199 to listen.
Pmgxamming
Example-Place
the Model 199 in remote
with the following statement:
PEMOTE 7215
The Model 199 should be in remote, as indicated
by the
annunciator
light. If not, check to see that proper bus con-
nections are made, and that the instrument
is programm-
ed for the correct primary address
(26).
Note that all front panel controls except LQCAL (and, of
course, POWER) are inoperative
while the inshument
is
in remote. You can restore normal front panel operation
by pressing
the LOCAL button.
3.8.2 IFC (Interface
Clear)
The FC command
is sent by the controller to place the
Model 199 in the talker and listener idle states. The~unit
will respond to the IFC command by cancelling
TALK or
LISTEN, if the instrument
was previously
placed in one
of those modes.
To send the IFC command,
the controller
need only set
the IFC line true for a minimum
of 1OO~sec.
Table 3-6. General Bus Commands and Associated BASIC Statements
BASIC 4.0~
Statement
Affect on Model
199
REMOTE 7
Goes into remote when next addressed.
IZIEORT 7
Goes into talker and listener
idle states.
LOCAL LljC:Kijl-lT ;
Front LOCAL key locked out.
LljCkL 726
Cancel remote.
C'LEAR 7
Returns to default conditions.
CLEAR 726
Returns to default conditions.
TF: I GGER 725.
Triggers reading in TZ and T3 modes.
3-10