BASIC DMM OPERATION
2.7.4 Line Frequency
The programmed
line frequency should match that of the
power line voltage in order for the instrume~nt to meet its
noise specifications
at 5%-d@
resolution
(line cycle in-
tegration is used at~S%-digit resolution). The line frequency
program can be used to check the programmed
line fre-
quency and set it to 50 or 60Hz.
Proceed
as follows
in order
to check
or set the line
frequency:
1. Press SHIFI
DMM SETUP and then NEXT repeatedly
until the line frequency message is displayed. For 6OHz,
the display will show:
FREQ=60Hz
2. For 50Hz, the display message
is:
FREQ=50Hz
3. Use uprange or downrange
to toggle to the desired @e-
quency, then press_NEXT to go on to the next program.
NOTES:
1. To change the default frequency setting, ftist select the
desired frequency and then use the save setup program
(or send Ll over the bus) to save the new frequency set-
ting. Cycling power, or sending SDC, DCL, or L1, over
the bus will not affect the programmed
line frequency.
2. Jf the line frequency
is changed but not saved with the
save setup progmm, sending SDC or DCL over the bus
will return the line frequency
to the default setting.
However, the reset program will not have any effect orI
the current frequency
setting, and sending LO over the
bus will not change the setting, but will save the new
frequency.
3. An "UNCAL!' error will default the IEEE-488 primary
address
to 26 and set the line frequency
to 6OHz.
2.7.5 Save Setup
The save setup program allows you to save current instru-
ment conditions.
These conditions
will then be assumed
upon power up, or after the instrument
receives the DCL
or SDC command
over the IEEE-488 bus.
The following
operating
parameters
are saved by this
program:
Function
Range
Resolution
Zero state (on/off)
Filter state (on/off)
AC dB state (onioffl
Multiplexer
(on/off)
Trigger delay
Reading
interval
IEEE488
primary
address
Line frequency
(50 or 60Hz)
In order to save an instrument
setup, proceed as follows:
1. Setup instrument operating conditions as desired, or use
the reset
program
(paragraph
2.78)
to save default
operating
conditions.
2. Press SHIFT
DMM SETUP and then NEXT until the
following
message
appears:
SAVE NO
3. Use uprange
or downrange
to toggle to the following
message:
SAVES YES
4. To'save instrument
setup conditions,
press NEXT. The
unit will save the op~erating states and then go on to the
next program.
NOTES:
l.To
exit the program
without
changing
the previous
default conditions,
press NEXT with the "SAVE NO"
message
displayed.
2. To return the instrument to the factory power up default
conditions,
use the reset program and then save those
conditions
using the save setup program.