Keithley 199 Instruction Manual page 59

System dmm scanner
Hide thumbs Also See for 199:
Table of Contents

Advertisement

BASIC DMM OPERATION
The basic test procedure
is as follows.
1. Press SHIl!T TRIG SETW
and program the unit for
the one-shot
trigger mode.
2. Using SCAN SETUP, select the Z-pole, %I
scan, and
ratio off modes.
3. Select
the ACV function,
then select a range~large
enough for the expected amplifier
output voltages.
4. Program a channel limit of 8 with the SCANNER key.
5. Press SHIFT dB to select the dB function.
6. Set the signal generator to the desired amplitude
and
mid-band
frequency
(for example, lkHz).
7. Press ZERO and then TRIGGER
to store the OdB
reference value. The display should now show O.OOdB
on channel
1.
8. Lower the generator
frequency
until the Model 199
displays
-3.OdB. The present generator frequency
is
the lower half-power,
or -3dB response
point.
9. Raise the generator
frequency
above the mid-band
point until the display again reads -3dB. -l'Jxe generator
frequency
now represents
the upper half-power,
or
-3dB
response
point of the amplifier.
10. Press TRIGGER
to advance the channel.
11.~ Repeat steps 7 through 10 for the remaining channels.
2.11.12 Low-level Measurement Considerations
The relay contacts of the Model 1992 Scanner Card have
low-thermal characteristics (clr;V
offset), allowing the card
to be used for low-level measurements.
The following
paragraphs
discuss
methods
to minimize
the effe&
oft
potential
error sources.
Thermoelectric
Potentials
Thermoelectric
potentials
(thermal EMFs) are small elec-
tric potentials
generated by differences
in temperature
at
the junctions
of dissimilar
metals. Such thermoelectric
potentials
can seriously
degrade low-level measwement
accuracy. For example, a copper-to-cop~per oxide junction
may generate
up to lOOO~WC, while a clean copper-to-
copper junction
will typically generate
only O&WC
or
less.
Ways to minimize the generation of thermoelectric
poten-
tials include:
L Use only copper wires for all input and output connec-
tions. If lugs are used, they should be crimped on (not
soldered),
and they should also be made of copper.
2. Keep all connecting
surfaces clean and free of oxides.
Wires and lugs should be carefully cleaned before be-
ing mated together.
3. Keep connecting
points
and junctions
at the same
temperature.
4. Protect all circuits and connecting
points from drafts.
Shielding
Shielding is important to keep noise out of low-level signal
p"hs.
To m inimize problems in these areas, all input and
output connections
to the scanner card should be made
using shielded
cable when measuring
low-level signals.
The shields should be connected
to signal LQ (not earth
ground) at the scanner card end for scanner input con-
nections,
and at the DMM end for scanner card output
connections.
Note that only one end of the shields should
be connected
to avoid possible ground loop problems;
the
other ends of the shields should be left floating.
2.11.13 Using the Scanner with Other
Instrumentation
Although
the scanner card is intended
for use primarily
with the Model 199 DMM, it can also be wed with other
instrmnentatio".
For -pie,
assume that the Model 1992
is to be used with a Keithley Model 181 Nanovoltmeter
to
make PV measurements
requiring a higher input resistance
than is available with the Model 199.
Typical connections
for this arrangement
are shown in
Figure 2-K'. Here, the scanner card outputs are connected
to the Model
181 mV input using
low-thermal
cables.
Likewise,' all scanner
inputs must be made with low-
thermal cables. Use copper wire and keep all connections
clean and free of oxidation.
Also, all signal paths should
be shielded
as discussed
above.
2-37

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents