APPENDIX B
HEWLETT-PACKARD
MODEL 85
The following program sends a command
string to then Model 199 fr&n an HP-85 computer
and
displays the instrument
reading string on the computer
CRT. The computer
must be equipped
with the HP82937 GPIB Interface
and an 110 ROM.
DIRECTIONS
1. Erom the front panel,
set the primary
address of the Model 199 to 26.
2. With the power off, connect the Model 199 to the HP82937A GPIB interface installed in the HP-85
computer.
3. Enter the lines in the program below, using the END LINE key after each line.
4. Press the HP-85 RUN key and type in the desired command
string at the command
prompt.
For example, to place the instrument
in ACV and autorange,
type in FIROX and press the END
LINE kev.
5. The
im&ument
reading
string
will
then
appear ~OIJ the
CRT
A
typical
display
is:
NACV+O,OOOCWE+O.
PROGRAM
COMMENTS
Dimension
strings.
Place 199 in remote.
Prompt for command.
~~Input command
string.
Address 199 to listen, send string.
Address 199 to talk, input reading.
Display reading string.
Repeat
NOTE: For conversion
to numeric variable,
change line 70 as follows:
B-5