Keithley 199 Instruction Manual page 8

System dmm scanner
Hide thumbs Also See for 199:
Table of Contents

Advertisement

Scanner Operation
r
POLE
A/T
;
Mode
A/V
L
RATIO
*L
L
2
POLE
4
POLE
MANUAL:
A!lows channel to be manuaJly closed
with SCANNER
key.
STEP: lnaements
one channel perintenral or trigger,
SCAN: Scans one set of channels per interval or trig-
ger (minimum time between channels).
ON: Channels 2 through 8 referenced to channel 1.
OFF
l Ratio operates only on
a fixed range.
l
Range changes will restart at channel 1.
l
In MANUAL, at least one reading must be taken on channel 1 before at-
tempting
to display ratio on channels Z-8.
ACTIVATING THE SCANNER
SCANNER
followed by number
(%3) activates the scanner.
- In MANUAL,
channel number
selects channel to be dosed.
l
In STEP or SCAN, number selects channel limits and starts scanning
process.
STOPPING THE SCANNER (opening au aam&)
SCANNER
followed by "0':
DETERMINING
SCANNING
INTERVAL
- scanning without s&ding interval
--Program
INTERVAL SELECT to OFF.
-Interval
is sl75msec.
depending
on other selected parameters.
--INTERVAL
OVERRUN message wi!J not be displayed.
l
Mienurn
Interval Tiie
Calculation
-Interval
time is the sum oE
1. (conversions
per channel)
x (conversion
time)
2.
Programmed
DELAY time per channel
3. Break-before-make
time (17msec)
4.
Auto range time (if used)
--In !XEP mode, c&l&d
time above is the INTERVAL setting.
-In
SCAN mode, (sum of l-4) x (number channels scanned) is the
minimum interval setting.
SCANNER OPERATION
NOTES
l
When using the sc.mner with STEP or SCAN switching, the DMM wi!J
takereading~oneachcharmel
ssifit
were in&one-shot
mode whether
pmgmmmed
to CONTlNUOUS
or ONE SHOT:
-Conversions
are automatically
synchronized
to channel
closures.
Channels
will not close in the middle of conversions.
-Any
programmed
DELAY is inserted between channel closure and
start of conversion
(with scanner disabled,
DELAY is inserted bet-
ween trigger and start of conversion).
-Readings
on each channel consist of multiple conversions
to fill
the
Running
Average
User Filter (30 conversions),
01 Running Average
Internal Filter when active (51hd only, varies by range and function;
see manual).
* Scanner switching is break-before-make.
The time required to change
channels is approximately
17rwc.
which includes break-before-make
relay time settling time.
l
When using the scanner with ACV or ACA, a DELAY time must be
programmed to accomcdate AC converter settling time (typically >lsec).
See Model 199 sp&fications.
*The scanner operates with the set of insirument
parameters
program-
med prior to stat of scanning. Chanpinp any of these parameters, range
function,
filter, etc., during scanning will restart the scanner at than-
~11.
Use autorange if range changes are required while scanning.
USING SCANNER WITH DATA STORE
l
Select desired scanner parameters
and interval.
l
Activate scanning
(SCANNER,
channel limit).
- Program data store size.
. TRIGGER vd, start data store and automatically restart scanning at than-
nel 1 synchronized
with data store location 1.
EXAMPLE' OF FAST SCANNING
1. Select 3V DC range and function,
4%-d@
resolution,
FLTR off.
2. Program ONE SHOT trigger.
3.
Select SCAN mode.
4.
Select SCANNER
'8" to set limit to 8.
5. Use TRIGGER
to initiate a scan of the set of eight channels.