Keithley 199 Instruction Manual page 42

System dmm scanner
Hide thumbs Also See for 199:
Table of Contents

Advertisement

2.8.2 Trigger Sources
For standard bench operation, there are two trigger sources
available:
front panel TRIGGER
button, and the EXTER-
NAL TRIGGER
INPUT jack. Upon power up both these
trigger sources will be enabled. Additional triggers include
IEEE-488
X, GET, and talk commands,
as discussed
in
paragraph
3.9.7.
NOTES:
1. TRIGGER
is always enabled regardless
of the selected
trigger source;
however, all front panel buttons will be
locked out when the unit is in remote.
2.~ Triggering the unit while it is still processing
a reading
from a previous
trigger will generate
the "TRIGGER
OVERRUN"
message.
2.8.3 Trigger Delay
The trigger delay period is the time from the trigger point
until the unit takes a reading. This delay period is also used
after each channel closure when using the scanner. For the
continuous
mode, the delay period affects only the fast
conversion;
however, with the one-shot
mode, the delay
period
affects
every conversion,
with the instrument
waiting
the programmed
delay time after each tr&ger
before taking a reading.
For example,
if you program a
300msec trigger delay, the unit will wait 3i%msec after each
trigger before taking a reading.
The allowable range for the trigger delay peliod is between
0 and 999.999 seconds in one millisecond
increments.
The
trigger delay can be programmed
with the TRIG SETUP
key, as described
below.
1. Press SHIFT TRIG SETUP and then NEXTand
note the
following message
is displayed
briefly:
DELAY=
000.000s
Here, we have assumed the factory default delay period
of Osec.
2.
Using the the data entry keys (O-9), key in the d#red
trigger delay period in the range of O-9999YYsec.
3. Once
the desired
delay value is programmed,
press
NEXT to go on to the interval selection (or press NEXT
once more to return to normal display).
2.8.4 Reading Interval
The reading interval parameter determines the time period
between individual readings when the instrument is in the
continuous
trigger mode. Interval also affects the rate of
data store operation, as discussed in paragraph 2.10, as well
as the interval between channels
(step mode) or scan se-
quences (scan mode) when using the optional Model 1992
Scanner
(see paragraph 2.11).
The unit can beprogrammed
for either default or selected
interval operation.
With default interval ("SELECT OFF'!),
a preset interval of l7Smsec is automatically selected. With
selected interval ("SELECT ON") a user-defined
interval
can be programmed.
The allowable range for selected in-
terval is l5msec
to 999.999sec
in lmsec increments.
NOTE
Programming
too short an interval for the present
instrument configuratiozi when using the scanner
or data store will result in the "INTERVAL OVER-
RUN" message. The interval cannot be programm-
ed if the unit is in the one-shot
trigger mode.
Reading interval can be checked or programmed
with the
TRIG SETUP key, as follows.
1. Press SHTn
TRIG SETUP and then NEXT repeatedly
until the following
is displayed:
INTERVAL
Next, the unit displays:
SELECT OFF
Or;~
SELECT ON
2. Use uprange or downrange
to select th-e-desired inter-
val type, select or default, then press NEXT. If default
(SELEa
OFF) is selected,
the interval will be set to
l75msec,
and the unit Will return to normal display.
3. If select interval is chosen, the presently selected
inter-
val will then be displayed,
as in this example:
INTERVAL =
ooo.l.75 s
Here we have assumed
the default interval of l75msec.
4. If desired,
key in a new reading interval fin the range
of 15msec to Y99.999seC.
5. Once the desired interval is displayed,
press NEXT to
return to normal display. If you progam too small =*
interval, the following will be displayed:
MIN = .OlS S
2-20

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents