Environmental Conditions; Warm-Upperiod; Cal Lock Switch; Front Panel Calibration - Keithley 199 Instruction Manual

System dmm scanner
Hide thumbs Also See for 199:
Table of Contents

Advertisement

MAINTENANCE
Table 6-4. Recommended Calibration Equipment
Mfg
Model
Description
Fluke
5440A
DC Voltage Calibrator
Fluke
52OOA AC Voltage Calit+or
Specifications
3OOmV, 3V, 3OV, 300V ranges
*l5ppm
33OmV, 3% 3OV ranges;
20Hz 20.1%;
50Hz-20kHz
0.02%: 1OOkH.z +0.33%
/ Fluke
1 52l5A 1 AC Power Atiplifiier
;/ 3oOV range; 20Hz iO.1246; 50i?z-2dkHz
/
&0.04%; iOOkHz +O.l%
Fluke
5450A
Resistance
Calibrator
3000-3MQ
ranges;
+l5ppm;
30MQ
+32ppm;
3OOMR @5ppm
Valhalla
2500E
Current Calibrator
30mA,~3A ranges
+O.OU%
6.4.2 Environmental
Conditions
Calibration should be performed under laboratory condi-
tions having an ambient temperature
of 23°C +l'C and
a relative humidity
of less than 70%.
6.4.3 Warm-Up Period
Turn on the instrument
power and allow it to warm up
for at least two hours before beginning the calibration pro-
cedure. If the instrument
has been subjected~to &r&es
of temperature
or humidity, allow at least one additional
hour for the instrument
to stabilize before beginning
the
calibration
procedure.
6.4.4 CAL LOCK Switch
Upon power up, calibration
is automatically
locked out.
A CAL LOCK switch,
accessible
from the front panel,
disables or enables front panel and IEEE488
bus calibra-
tion. When calibration
is disabled,
calibration
cannot be
performed. The following. message will be briefly displayed
when attempting
to enter the calibration
p+gram
while
the switch is disabled:
CAL LOCKED
Calibration
can only be accomplished
with calibration
enabled.
To enable calibration,
push in on the CAL LOCK switch
with an indated
tool (through the smaU access hole under
the round label). The instrument
will briefly display the
following message:
CAL UNLOCK
After calibration is complete, press in on CAL LOCK a se-
cond time in order to disable calibration. The unit will then
display the "CAL LOCKED"
message.
NOTE
The two detiinal points in the function area of the
display will be on when calibration
is enabled.
6.4.5 Front Panel Calibration
The following
information
provides the basic procedure
for calibrating
the instrument
from ~the front panel.
A
detailed calibration procedure is located in paragraph 6.4.7
NOTE
NOTE
DCV ranges
have three
cal points:
minus
full
DCV ranges
have three
cal points:
minus
full
range, plus full range, and zero. The remaining
range, plus full range, and zero. The remaining
functi&s
require only two cal points.
functi&s
require only two cal points.
1. Place the tinit fin the "CAL UNLOCK"
state to enable
calibration. The switch is accessed from the front panel
of the instrument
through the access hole.
2. Select
the function
and range to be calibrated
(DC
VOLTS, AC VOLTS, OHMS, DC AMPS, or AC AMPS),
5% digit resolution,
zero off, and filter off.
3. Select the front panel calibration
program as follows:
A. Press SHW
LOCAL. The following message will
be displayed
briefly:
CAL=
B. The default calibration
point, which is a high end
reading for the selected
range and function,
will
now be displayed.
For example, if the 3V DC range
was selected
in step 2, the following
calibration
point will be displayed:
6-3