BASIC DMM OPERATION
press the NEXT key. The unit will then display the data
value at that location along with the measurement
func-
tion in effect ate the time the data was taken.
5.
To exit the recall mode, press NEXT while normal recall
data is displayed. RCL Will turn off ~to indicate that recall
has been disabled.
NOTES:
1. If data store has no valid data to display, the unit will
display the following
message upon entry to the recall
mode.
NO DATA
2.~The unit will continue
to store data while in recall until
the data store buffer is full (or continuously
in wrap
around mode).
2.11 SCANNER OPERATION
(WITH OPTION 1992)
With the optional Model 1992 Scanner installed, the Model
199 &m scan four, 4-pole channels,
or eight, 2-pole chan-
nels. The following paragraphs
discuss scanner program-
ming, connections,
atid operation
from the front panel.
Refer to paragraph
3.12 for IEEE-488 scar&r
programm-
ing. For scanner installation procedures,
refer tom Section 6.
2.11.1 Scanner Connections
Figure 2-10 shows the Model 1992 Scanner
Card and its
two quick-disconnect
terminal
blocks.
To remove
each
block from the card, simply pull on the attached
handle
until it comes free of the card. Screw terminals
on the
blocks accept up to #14AWG solid or stranded
wire.
1992 CARD
I"
Figure 2-10. Scanner Connections