Sdc (Selective Device Clear); Get (Group Execute Trigger) - Keithley 199 Instruction Manual

System dmm scanner
Hide thumbs Also See for 199:
Table of Contents

Advertisement

lEEE-488
PROGRAMMING
To send the DCL command,
the controller must perform
the following
steps:
1. Set ATN true.
2. Place the DCL command
byte on the data bus.
Notes:
1. DCL will return the instrument
to the default line fre-
quency setting.
2. DCL will not have any effect
on the current
IEEE
address.
Programming
Example--Place
the unit in an operating
mode that is not a default condition. Now enter the follow-
ing statement
into the keyboard:
Note that the instrument returns to the default conditions.
3.8.6 SDC (Selective
Device Clear)
The SDC command
is an addressed
command
that per-
forms essentially the same function as the DCL command.
However, since each device must be individually
address-
ed, the SDC command
provides a method to clear only
a single,
selected
instrument
instead
of clearing
all in-
struments simultaneously,
as is
the
case with DCL.-When
the
Model
199 receives the SDC command,
it wil1 return
to the default conditions
(see paragraph
3.9.11). Table 37
lists factory default conditions.
To transmit the SDC command,
the controller
must per-
form the following
steps:
1. Set ATN true.
2. Address the Model 199 to listen.
3. Place the SDC command
byte on the data bus.
Notes:
1. SDC~ will return the instrument
to the default line fre-
quency setting.
i.
<DC
will not have any effect on the current
IEEE
address.
Programming
Example-Using
several front panel con-
trols, alter instrument
states from the default configura-
tion
Send SDC with the following
statement:
When the above statement
is executed,
the instrument
returns to the default configuration.
3.8.7 GET (Group Execute Trigger)
GET
may
be used to initiate a Model 199 measurement
se-
quence if the instrument
is placed in the appropriate
trig-
ger mode (see paragraph 3.9). Once triggered, the instru-
ment will take a single reading or series of readings.
To send GET
the controller
must perform the following
sequence:-
1. Set ATN low.
2. Address the Model 199 to listen.
3. Place the GET command
byte on the data bus.
Programming
Example-Type
in the following statements
to place the instrument in the correct hitzger mode for pur-
poses of this demonstration:
Now trigger the reading by sending GET with the follow-
ing statement:
The
reading
will be triggered
when
the statement
is
executed.
3-12