The adjusted measurement values appear after probe reset.
Performing analog adjustment
Please refer to chapter 1.3.5.4, page 63.
In Measuring Mode, press SET, select Main Menu Adjust using
1
or and confirm selection with SET.
Each channel is adjusted at three points in the analog range
(at 10 %; 50 %; 90°% of the analog scale).
Select Analog Adj. Ch. 1 using or and confirm with SET.
2
3
Select Adj. Point 1 using or .
4
Press SET. Read off multimeter display (e.g. 5.601 mA) and enter
this value in the user menu. Do this by scrolling one digit to the
right using and increasing the value of digit by 1 using .
Confirm with SET or abort entry with ESC.
5
Select Adj. Point 2 using or .
6
Press SET. Read off multimeter display (e.g. 12.001 mA) and
enter this value in the user menu. Do this by scrolling one digit to
the right using and increasing the value of digit by 1 using .
Confirm with SET or abort entry with ESC.
7
Select Adj. Point 3 using or .
8
Press SET. Read off multimeter display (e.g. 18.401 mA) and
enter this value in the user menu. Do this by scrolling one digit to
the right using and increasing the value of digit by 1 using .
Confirm with SET or abort entry with ESC.
Continue to Analog Adj. Ch. 2 or 3 with or (repeat steps 3 to
9
8).
10 Return to Main Menu Adjust by pressing ESC.
11 Continue to Main Menu Reset using or or return to
Measuring Mode by pressing ESC.
testo 6681 Ethernet
- 1.4 Operation 85