TESTO 6681 Instruction Manual page 34

Humidity transmitter with ethernet module
Hide thumbs Also See for 6681:
Table of Contents

Advertisement

testo 6681 Ethernet
34
Using the earthing terminal within the instrument
1
Guide PE line (yellow-green) (5) through the cable coupling (x)
and fit cable lug (8). Fix this to the side of the instrument (6) using
M 5 screw (3), washer (4) and snap ring (7) on the internal
earthing terminal (1).
2
Place the other cable end on an appropriate (PE) earth conductor,
e.g. an earthing bar.
Using an earthing terminal outside of the instrument
1
Use shielded cable (5) with cable lug (8). Fix this in place using
M 5 screw (3), washer (4) and snap ring(7) on the external
earthing terminal (2).
2
Place the other cable end on an appropriate (PE) earth conductor,
e.g. an earthing bar.
1.3.3.6
Setting the Ethernet module
Using a DIP switch the Ethernet module can be switched between two main
functions:
Saveris subscriber function (DIP switch No 1 = on)
testo 6681 is used as a testo Saveris component.
XML server function (DIP switch No. 1 = off)
integration of the testo 6681 into the customer's Ethernet system.
The testo 6681 must be disconnected from the Ethernet network.
- 1.3 Commissioning
Important:
On delivery, the testo 6681 with Ethernet module is in Saveris
mode (switch setting "on").