Using ESC return to H
Absolute pressure unit using or .
Selecting absolute pressure unit (Abs. pressure unit)
This parameter determines the humidity variables, standardised atmospheric
dewpoint (°CtA, °FtA), relative humidity (g/kg or gr/lb) and water content
(ppm
or % vol).
vol
1
Press SET, select desired unit (selection: bar, psi, mPa, hPa)
using or confirm selection with SET or cancel with ESC.
Continue to Absolute Pressure with or .
2
Editing absolute pressure (Abs. pressure value)
You can set a value for the process absolute pressure.
1
Using or , select Abs. pressure value and confirm selection
with SET.
The absolute pressure is displayed.
2
Scroll one digit to the right using , increase value of digit by 1
using . Confirm with SET or abort entry with ESC.
3
Press SET and return to Abs. pressure value.
4
Return to Main Menu Settings by pressing ESC.
Continue to Main Menu Analysis using or or return to
5
Measuring Mode by pressing ESC.
Editing code settings
You can set the access code (password).
If a code other than "0000" (factory setting) is set, the
transmitter can only be operated once this code has been
entered via the menu.
Select Code using or and confirm selection with SET.
1
2
Scroll one digit to the right using , increase value of digit by 1
using . Confirm with SET or abort entry with ESC.
3
Return to Code by pressing ESC.