TESTO 6681 Instruction Manual page 67

Humidity transmitter with ethernet module
Hide thumbs Also See for 6681:
Table of Contents

Advertisement

The process temperature should not vary by more than 0.5K
The dew point temperature should remain stable, as far as possible
The process pressure should not vary excessively
If these prerequisites cannot be fulfilled, the values obtained
during the last successful self-adjustment will be maintained.
All adjustment processes are saved in the history of the transmitter, see
Transmitter history, page 172.
testo 6615:
testo 6615:
probe
Fühler-
behaviour
verhalten
Analog
Analog-
output °Ctpd/
Ausgang
°Ftpd
°Ctpd/
°Ftpd
The heating time and storing of cycles can be edited in the P2A software; for
example, they can be deactivated by setting the two parameters to "0".
testo 6681 Ethernet
Parameter
Parameter
"Duration self-adjustment time""
„Dauer Heizphase"
"Self-adjustment cycle time"
„Zykluszeit Selbstabgleich"
- 1.3 Commissioning 67

Hide quick links:

Advertisement

Table of Contents
loading