Chapter 5 To Trigger the Oscilloscope

Delay Trigger (Option)

In delay trigger, you need to set signal source A and signal source B. The oscilloscope
triggers when the time difference (△T) between the specified edges of source A
(Edge A) and source B (Edge B) meets the preset time limit, as shown in the figure
below.
Note:
Edge A and Edge B must be neighbouring edges.
Source B
Trigger Type:
Press Type to open the trigger type list. Rotate
. At this point, the trigger setting information is displayed at the upper right corner
of the screen.
For example,
trigger source is CH1; the trigger level is 0.00 V.
Source A:
Press SourceA to select CH1-CH4 as the trigger source of signal source A.
Edge A:
Press EdgeA to select the trigger edge type of signal source A in delay trigger. It can
be set to the rising edge or falling edge.
Source B:
Press SourceB to select CH1-CH4 as the trigger source of signal source B.
Edge B:
Press EdgeB to select the trigger edge type of signal source B in delay trigger. It can
be set to the rising edge or falling edge.
Delay Type:
Press DelayType to set the time limit condition of delay trigger.
MSO1000Z/DS1000Z User's Guide
Source A
T
Figure 5-10 Delay Trigger
. The trigger type is delay trigger; the current
Edge A=
Edge B=
to select "Delay" and press down