RIGOL
occurs, the oscilloscope will trigger on the read/write bit. After this trigger
condition is selected:
--press AddrBits to select "7 bits", "8 bits" or "10 bits";
--press Address to set the address value of I2C trigger. According to the setting
in AddrBits, the range can be from 0 to 0x7F, from 0 to 0xFF or from 0 to
0x3FF;
--press Direction to select "Read", "Write" or "R/W".
available when AddrBits is set to "8 bits".
Data: the trigger searches for the specified data value on the data line (SDA).
When this event occurs, the oscilloscope will trigger on the clock line (SCL)
transition edge of the last bit of data. After this trigger condition is selected:
--press Bit X to select the desired data bit and the range is from 0 to (Byte
Length×8-1);
--press Data to set the data pattern of the current data bit to X, H or L;
--press Bytes to set the length of the data and the range is from 1 to 5;
--press AllBits to set the data pattern of all the data bits to the data pattern
specified in Data.
A&D: the oscilloscope searches for the specified address and data at the same
time and triggers when the specified address and data are found. After this
trigger condition is selected:
--press AddrBits to select "7 bits", "8 bits" or "10 bits";
--press Address and use
in AddrBits, the range can be from 0 to 0x7F, from 0 to 0xFF or from 0 to
0x3FF;
--press Bit X to select the desired data bit and the range is from 0 to (Byte
Length×8-1);
--press Data to set the data pattern of the current data bit to X, H or L;
--press Bytes to set the length of the data and the range is from 1 to 5;
--press AllBits to set the data pattern of all the data bits to the data pattern
specified in Data;
--press Direction to select "Read", "Write" or "R/W".
Note:
This setting is not available when AddrBits is set to "8 bits".
Trigger Mode:
Press Sweep to open the trigger mode list and select auto, normal or single. For the
details, please refer to "Trigger Mode". The corresponding status light of the
current trigger mode turns on.
Trigger Setting:
Press Setting to set the trigger parameter (noise rejection) under this trigger type.
Trigger Level:
When SCL is an analog channel, press SCL and use TRIGGER
the trigger level of the SCL channel. When SDA is an analog channel, press SDA and
use TRIGGER
details, please refer to the introduction of "Trigger Level".
5-34
to set the address value. According to the setting
LEVEL to modify the trigger level of the SDA channel. For the
Chapter 5 To Trigger the Oscilloscope
Note:
This setting is not
LEVEL to modify
MSO1000Z/DS1000Z User's Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds1000z series

Table of Contents