Rigol MSO1000Z series User Manual page 133

Hide thumbs Also See for MSO1000Z series:
Table of Contents

Advertisement

Chapter 6 MATH and Measurement
Delay and Phase
Source 1
Source 2
Source 1 and source 2, the same as source A and source B in the measurement
setting menu can be any channel of CH1-CH4 or D0-D15. For the setting method,
please refer to the introduction in "Measurement Setting".
1. Delay 1→2 : the time difference between the rising edges of source 1 and
source 2. Negative delay indicates that the selected rising edge of source 1
occurred after that of source 2.
2. Delay 1→2 : the time difference between the falling edges of source 1 and
source 2. Negative delay indicates that the selected falling edge of source 1
occurred after that of source 2.
3. Phase 1→2 : phase difference calculated according to "Delay 1→2 " and the
period of source 1, expressed in degree. The calculation formula is as shown in
equation (6-1).
4. Phase 1→2 : phase difference calculated according to "Delay 1→2 " and the
period of source 1, expressed in degree. The calculation formula is as shown in
equation (6-1).
The phase calculation formula:
Wherein,
represents "Phase 1→2 " or "Phase 1→2 "
Phase
represents "Delay 1→2 " or "Delay 1→2 "
Delay
represents the period of source 1
Period
1
MSO1000Z/DS1000Z User's Guide
Period
Delay
Figure 6-4 Delay and Phase
Delay
Phase =
Period1
°
×360
RIGOL
(6-1)