Function Inspection - Rigol MSO1000Z series User Manual

Hide thumbs Also See for MSO1000Z series:
Table of Contents

Advertisement

Chapter 1 Quick Start

Function Inspection

Press Storage  Default to restore the oscilloscope to its default
1.
configuratiuon.
2.
Connect the earth alligator clip of the probe to the "Ground Terminal" as shown
in the figure below.
3.
Use the probe to connect the input terminal of CH1 of the oscilloscope and the
"Compensation Signal Output Terminal" of the probe.
Press AUTO.
4.
5.
Observe the waveform on the display. In normal condition, the square waveform
as shown in the figure below should be displayed.
6.
Use the same method to test the other channels. If the square waveforms
actually shown do not match that in the figure above, please perform "Probe
Compensation".
WARNING
To avoid electric shock when using the probe, please make sure that the
insulated wire of the probe is in good condition and do not touch the
metallic part of the probe when the probe is connected to high voltage
source.
Tip
The signal output from the probe compensation connector can only be used for
probe compensation adjustment and can not be used for calibration.
MSO1000Z/DS1000Z User's Guide
Figure 1-7 To Use the Compensation Signal
Figure 1-8 Square Waveform
Compensation Signal Output Terminal
Ground Terminal
RIGOL
1-7