11.5.6 Numbering of measuring operations

To identify measuring operations or series of measuring operations these can
be individually numbered before starting. This number is output or saved when
the next measuring point scan starts. In this way individual measuring opera-
tions can be assigned to certain types of measurement or certain measuring
points (see Manual 6.7).
After selecting the function
(see 8.5). You can use digits 0 to 9 and also the characters A, F, N, P, and - or
_ (space). The number is activated as soon as it has been entered; it will then
be followed by the letter ´A´ until the next cyclic or manual measuring operation
is saved.
Number function : (e.g. room 12, measuring point 1)
To zero-set and deactivate the number press
To activate and deactivate the number press :
To increment and activate the number press

11.5.7 Memory space, memory output, clearing the memory

When measured values are being recorded the
tion continuously displays the memory capacity still available. Selecting this
function enables two softkeys, one for direct memory output and one for mem-
ory clearing. The output format is as set in the cycle (see 11.5.5, 11.5.8.1, and
Manual 6.6.1).
Function
Memory free
To output the memory :
To clear the memory :
Using the keypad the contents of the measured value memory can only be out-
put to the serial interface as a complete whole. Only by software can certain
sections of the memory be specified for output - either by stipulating the start
time and end time or by selecting the numbers defining the measuring opera-
tions in question. With each output from the internal memory any one of the
three output formats may be used "list", "columns", or "table".
With external MMC memory cards (see 11.5.2) the device itself can only read
out in table mode the measured data contained in the file most recently used.
The most sensible approach is to remove the memory card and copy the files
via a USB card reader directly onto the PC. These can then be imported either
into MS-Excel or into Win-Control (as of V.4.8.1).
During memory output the
and displays in KB the amount still to be output.
Remaining amount of memory to be output
the 6-digit number is entered as normal
Number
e.g.:
Remaining output function
Data logger functions
Number:
<CLR>
,
<ON>
<OFF>
< +1 >
Memory capacity free
Memory free :
<PMEM>
<CMEM>
continuously updates
Remaining output :
ALMEMO
®
2590
12-001
A
func-
38.4
KB
12.5 KB
41