753/754
Users Manual
Simultaneous Measure/Source
Use the MEASURE/SOURCE mode to calibrate or emulate
a process instrument. Push so that a split screen
display shows as in Figure 27.
Figure 27. Measure and Source Screen
66
info@Fluke-Direct.ca
Table 9 shows the functions that can be used at the same
time when Loop Power is disabled. Table 10 shows the
functions that can be used at the same time when Loop
Power is enabled.
Step or Auto Step features can be used to adjust the
output in MEASURE/SOURCE mode, or use the calibration
routine given when the As Found softkey is pushed.
Use the two softkeys shown in MEASURE/SOURCE mode
when you calibrate a process instrument:
•
As Found, which can be used to set up a calibration
routine to get and record as found data.
•
Auto Step, which can be used to set up the Product for
auto-stepping, as given before.
gks42s.bmp