Switch Calibration

The procedure to calibrate a switch also uses the As Found
and As Left calibration templates. Select the 1 Pt. Switch or
2 Pt. Switch procedure in a menu after you push the As
Found softkey. Figure 35 specifies the terminology used
when you calibrate limit switches.
The template to set up the switch procedure uses these
parameters:
Switch sense (normally open or closed)
For each setpoint:
Setpoint value
Setpoint tolerance
High limit or low limit
Minimum deadband
Maximum deadband
76
info@Fluke-Direct.ca
Process
Variable
Switch A: Low Limit
Figure 35. Switch Terminology
Fluke-Direct
.ca
Switch B
Open
Switch B
Setpoint
Deadband
Reset
Switch B
Closed
Switch A
Closed
Reset
Deadband
Switch A
Setpoint
Switch A
Open
Switch B: High Limit
1.800.868.7495
gj24f.eps