Select the √ instrument procedure in a menu after you push
the As Found softkey.
Switch Calibration
The procedure to calibrate a switch also uses the As Found
and As Left calibration templates. Select the 1 Pt. Switch or
2 Pt. Switch procedure in a menu after you push the As
Found softkey. Figure 35 specifies the terminology used
when you calibrate limit switches.
The template to set up the switch procedure uses these
parameters:
•
Switch sense (normally open or closed)
•
For each setpoint:
•
Setpoint value
•
Setpoint tolerance
•
High limit or low limit
•
Minimum deadband
•
Maximum deadband