753/754
Users Manual

RTD Simulation

Note
See Table 8 for data about RTD (Resistance-
Temperature Detector) types compatible with the
Product.
Connect the Product to the instrument under test as shown
in Figure 23. The figure shows connections for two, three, or
four-wire transmitters. For three or four-wire transmitters,
use the 4-inch long stackable jumper cables to connect the
third and fourth wires at the source V Ω RTD jacks.
To simulate an RTD (Resistance-Temperature Detector):
1.
If necessary, push  for SOURCE mode.
2.
Push t.
3.
Push  or  to select RTD.
4.
Push . The Select RTD Type display is shown.
5.
Push  or  followed by  to select the necessary
RTD type.
6.
The product tells you to use the keypad to enter the
temperature to simulate. Input the temperature, and
then push .
56