753/754
Users Manual
RTD Simulation
Note
See Table 8 for data about RTD (Resistance-
Temperature Detector) types compatible with the
Product.
Connect the Product to the instrument under test as shown
in Figure 23. The figure shows connections for two, three, or
four-wire transmitters. For three or four-wire transmitters,
use the 4-inch long stackable jumper cables to connect the
third and fourth wires at the source V Ω RTD jacks.
To simulate an RTD (Resistance-Temperature Detector):
1.
If necessary, push for SOURCE mode.
2.
Push t.
3.
Push or to select RTD.
4.
Push . The Select RTD Type display is shown.
5.
Push or followed by to select the necessary
RTD type.
6.
The product tells you to use the keypad to enter the
temperature to simulate. Input the temperature, and
then push .
56