753/754
Users Manual

Temperature, Resistance Temperature Detectors

Range °C
Type (α)
-200 to 100
0.07 °C
100 Ω
Pt(385)
100 to 800
0.02 % + 0.05 °C
-200 to 100
0.07 °C
200 Ω
Pt(385)
100 to 630
0.02 % + 0.05 °C
-200 to 100
0.07 °C
500 Ω
Pt(385)
100 to 630
0.02 % + 0.05 °C
-200 to 100
0.07 °C
1000 Ω
Pt(385)
100 to 630
0.02 % + 0.05 °C
-200 to 100
0.07 °C
100 Ω
Pt(3916)
100 to 630
0.02 % + 0.05 °C
-200 to 100
0.08 °C
100 Ω
Pt(3926)
100 to 630
0.02 % + 0.06 °C
116
info@Fluke-Direct.ca
Temperature, RTDs
Degrees or % of Reading
[2]
Measure °C
1 Year
2 Year
0.14 °C
0.04 % + 0.10 °C
0.14 °C
0.04 % + 0.10 °C
0.14 °C
0.04 % + 0.10 °C
0.14 °C
0.04 % + 0.10 °C
0.14 °C
0.04 % + 0.10 °C
0.16 °C
0.04 % + 0.12 °C
Fluke-Direct
[1]
Source °C
Source
1 Year
Current
0.05 °C
1 mA
0.0125 % + 0.04 °C
0.10 °C
500 μA
0.017 % + 0.09 °C
0.08 °C
250 μA
0.017 % + 0.06 °C
150 μA
0.06 °C
0.017 % + 0.05 °C
0.05 °C
1 mA
0.0125 % + 0.04 °C
0.05 °C
1 mA
0.0125 % + 0.04 °C
.ca
1.800.868.7495
Allowable
Excitation
2 Year
[3]
Current
0.10 °C
0.1 to 10 mA
0.025 % + 0.08 °C
0.20 °C
0.1 to 1 mA
0.034 % + 0.18 °C
0.16 °C
0.1 to 1 mA
0.034 % + 0.12 °C
0.12 °C
0.1 to 1 mA
0.034 % + 0.10 °C
0.10 °C
0.1 to 10 mA
0.025 % + 0.08 °C
0.10 °C
0.1 to 10 mA
0.025 % + 0.08 °C