Source Pressure

The Product has a source pressure display function where
an external pressure hand pump is necessary. Use this
function to calibrate instruments where a pressure source or
differential pressure measurement is necessary. See
Figures 21 and 36 for information about that application.
Many ranges and types of pressure modules are available
from Fluke, see "Accessories". Before you use a pressure
module, read its instruction sheet. The modules are different
in how they are used, zeroed, what types of process
pressure media are allowed, and accuracy specifications.
To use the source pressure display, see Figure 21:
Warning
To avoid a violent release of pressure in a
pressurized system, shut off the valve and
slowly bleed off the pressure before attaching
the pressure module to the pressure line.
52
info@Fluke-Direct.ca
To avoid mechanically damaging the pressure
module:
Never apply more than 10 ft.-lb. of torque
between the pressure module fittings or
between the fittings and the body of the
module.
Always apply correct torque between the
pressure module fitting and connecting
fittings or adapters.
To avoid damaging the pressure module
from overpressure, never apply pressure
above the rated maximum printed on the
pressure module.
To avoid damaging the pressure module
from corrosion, use it only with specified
materials. See the printing on the pressure
module or the pressure module instruction
sheet for the acceptable material
compatibility.