Steps 8 Through 11: 100 Ohms Measure, Resistance Measure Adjustment - Fluke 7526A-156 Calibration Manual

Temperature, pressure and low dc voltage precision process calibrator
Table of Contents

Advertisement

Steps 8 through 11: 100 Ohms Measure, Resistance Measure Adjustment

The ohms measure calibration adjustment allows the use of a medium-quality resistor
decade in conjunction with a good-quality DMM by using flexible calibration points.
This way, if resistance standards are not available, the resistor decade can be set to 1 K,
the actual value of the DMM value can be read (1002.45 Ω for example), and this value
can be entered when asked for 1000 Ω.
You can use a resistor decade box, adjustable in 25 Ω to 4000 Ω range, to cover all of the
values that are requested (maximum 5 % tolerance):
25 Ω
50 Ω
100 Ω
200 Ω
400 Ω
1000 Ω
2000 Ω
4000 Ω
The subsequent procedure uses the Fluke 5522 for ohms source to adjust the Product
reading. See Table 3 for the test equipment used for this procedure. See Figure 15 for
connections.
Calibrator
5522A
CALIBRATOR
Figure 15. 100 Ohms Measure, Resistance Measurement Adjustment Connections
STBY
OPR
7
8
4
5
1
2
/ +
0
Precision Process Calibrator
Calibration Adjustment
7526A
7526A
PRECISION PROCESS CALIBRATO
VOLTS
mA
RTD/
100 mA MAX
OUTPUT
HI
HI
OUTPUT
100V MAX
LO
LO
20 V PK
MAX
4 W RTD/
INPUT
HI
CURRENT
SENSE
LO
20 V PK
MAX
gxm024.eps