Summary of Contents for Fluke 753

  • Page 1 753/754 Documenting Process Calibrator Users Manual July 2011 Rev. 1, 8/12 © 2011-2012 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 2 LIMITED WARRANTY AND LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling.
  • Page 3: Table Of Contents

    Table of Contents Title Page Introduction ........................1 How to Contact Fluke ..................... 1 Safety Information ......................2 Standard Equipment....................... 6 Functions ........................9 Get Started ........................11 Operation Features ......................13 Input and Output Jacks ....................13 Buttons ........................15 Display ........................
  • Page 4 753/754 Users Manual Preserve Battery Life ....................25 The Battery Charger ....................25 Display Languages ......................26 Display Intensity......................26 Date and Time ....................... 26 The Backlight ......................... 28 Personalize the Product ....................28 Measure Mode ....................... 29 Measurement Ranges ....................29 Electrical Parameter Measurement ................
  • Page 5 The Calculator ........................ 87 Save to and Recall from the Registers ..............88 Use the Calculator to Set the Source Value .............. 88 Quick Guide to Applications ................... 88 Communication with a PC ....................101 Maintenance ........................101 Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 6 753/754 Users Manual Battery Replacement ....................101 Clean the Product ..................... 101 Calibration Data ......................102 In Case of Difficulty ....................102 Service Center Calibration or Repair ................ 102 User-Replaceable Parts ....................102 Accessories ........................104 Specifications ........................ 106 General Specifications ....................106 Environmental Specifications ..................
  • Page 7 Typical Battery Life ........................ 23 Thermocouple Types Accepted ..................... 37 RTD Types Accepted ......................39 Simultaneous MEASURE/SOURCE Functions with Loop Power Disabled ......67 Simultaneous MEASURE/SOURCE Functions with Loop Power Enabled ......68 Duration Limits ........................84 Replacement Parts ........................ 103 Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 8 753/754 Users Manual Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 9 Time and Date Display ......................27 Edit the Date Format ......................27 Personalize the Product ......................28 Electrical Measurement Connections ..................31 Gage and Differential Pressure Modules ................33 Pressure Measurement Connections ..................35 Temperature Measurement with a Thermocouple ..............38 Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 10 753/754 Users Manual Correct Jumper Use ......................41 Temperature Measurement with an RTD ................42 Electrical Source Connections ....................47 Connections fo Simulate a 4 to 20 mA Transmitter ............... 49 Connections to Supply Loop Power ..................51 Connections to Source Pressure ..................54 Connections to Simulate a Thermocouple ................
  • Page 11 Analog and HART Pressure Transmitter Connection ............95 mV to Current Transmitter Calibration ................... 96 Vortex Sheding Flowmeter Check ..................97 HART and Analog RTD Transmitter Connections ..............98 Analog and HART Thermocouple Transmitter Connections ..........99 Transmitter HART- Comm Only .................... 100 Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 12 753/754 Users Manual Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 13: Introduction

    Introduction How to Contact Fluke The 753 and 754 Documenting Process Calibrators (the To contact Fluke, call one of the following telephone Product) are battery-powered, hand-held instruments that numbers: measure and source electrical and physical parameters. In • Technical Support USA: 1-800-44-FLUKE ...
  • Page 14: Safety Information

    753/754 Users Manual • Safety Information Limit operation to the specified measurement category, voltage, or A Warning identifies conditions and actions that pose amperage ratings. hazards to the user; a Caution identifies conditions and • actions that may damage the Product or the equipment Do not exceed the Measurement Category under test.
  • Page 15 Carefully look at the insulation to the current terminals. around the terminals. • Use only cables with correct voltage • Always put the stackable end of the test ratings. lead into a terminal of the Product. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 16: Symbols

    WEEE shock. Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation” product. Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 17   Double insulated German certifying body. CAT II equipment is designed to protect against transients from energy-consuming equipment supplied from CAT II the fixed installation, such as TVs, PCs, portable tools, and other household appliances. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 18: Standard Equipment

    Battery with integrated charger/power supply and HART communications cable (754) international adapters • Calibration Manual (available from Fluke’s website) • Printed multilingual 753/754 Getting Started Manual • Sample DPCTrack2 application software • 753/754 Manual CD containing multilingual Users • NIST-traceable Certificate of Calibration Manuals •...
  • Page 19 Documenting Process Calibrator Standard Equipment TC Cap AC280 SureGrip Hook Clip Set (2 Red and 2 Black) Jumper Fluke-75X-Stackable Test Lead Set (3 Red and 3 Black) 754-8016, Alligator Clip Set (3 Red and 3 Black) Strap TP220-1 Test Probe (3 Red and 3 Black) gks01f.eps...
  • Page 20 753/754 Users Manual Battery Charger/Universal Power Supply USB Cable Lithium Ion Battery Pack 753/754 Getting Started DPCTrack2 HART Communication Cable 753/754 Manuals gks02f.eps Figure 1. Standard Equipment (cont) Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 21: Functions

    • Automatic calibration procedures for transmitters and from one ramp increment to the next. limit switches when you use split screen For performance testing and calibration instructions, MEASURE/SOURCE mode. download the 753/754 Calibration Manual from Fluke’s website. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 22: Summary Of Source And Measure Functions

    753/754 Users Manual Table 2. Summary of Source and Measure Functions Function Measure Source  0 V to ±300 V 0 V to ±15 V (10 mA max) Volts dc Volts ac 0.27 V to 300 V rms, 40 Hz to 500 Hz No sourcing 0.1 V to 30 V p-p sine wave, or 15 V peak...
  • Page 23: Get Started

    (if the battery is outside of the Product, charge for display as shown in Figure 3. 5 hours). For more information, see “The Battery”. The Battery will only charge if the Product is off. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 24 753/754 Users Manual gks03f.eps Figure 2. Jumper Connections gks04s.bmp Figure 3. Measure/Source Example Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 25: Operation Features

    Jack for the battery charger/universal power supply (referred to as the battery Battery Charger jack charger throughout this manual). Use the battery charger for bench-top  applications where ac line power is available. USB port (Type 2) Connects the Product to a USB port on a PC.  Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 26 753/754 Users Manual gks05f.eps Figure 4. Input/Output Jacks and Connectors Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 27: Buttons

    Softkey functions are defined by the labels that show above the softkey during operation. Softkey DOCUMENTING PROCESS CALIBRATOR labels and other display text are shown in this manual in bold type, for example, Choices. gks06f.eps Figure 5. Buttons Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 28: Buttons

    Enters and exits Setup mode to change operating parameters.  (754) (754) Toggles between HART communication mode and analog operation. In calculator mode, this key supplies the square root function.  (753)  (753) Adjusts the Range of the Product. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 29 Toggles between ac voltage and frequency functions in MEASURE mode, or selects  frequency output in SOURCE mode. Numeric Used when a numeric entry is necessary.  keypad  Cycles the Product through MEASURE, SOURCE, and MEASURE/SOURCE modes.  Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 30: Display

    753/754 Users Manual • Measured Value: Shows the measured value in a Display Figure 6 and Table 5 show a typical display. The display selection of engineering units or percent of scale. shown is MEASURE mode. Near the top of the display is •...
  • Page 31 Documenting Process Calibrator Operation Features gks07c.eps Figure 6. Elements of a Typical Display Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 32: Elements Of A Typical Display

    753/754 Users Manual Table 5. Elements of a Typical Display Item Description Time and Date Display  HART Indicator  Loop Power Indicator  Battery Gauge  Backlight Indicator  Source Status  Undamped (Unsettled) Indicator  Custom Units Indicator ...
  • Page 33: Strap And Stand

    Product on any sturdy support. Figure 7 also shows how to open the Stand to put the Product at an optimal sight angle for bench top use. gks8f.eps Figure 7. Stand Use and Strap Attachment Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 34: The Battery

    Connect the battery charger to the mains battery. power outlet before the Product. • To charge the battery while it is inside the Product: Use only Fluke approved power adapters to charge the battery. Turn the Product OFF. • Keep cells and battery packs clean and Connect the battery charger to the Product and keep it dry.
  • Page 35: Battery Charge Level

    Battery Charge Indicator. LEDs flash to show the charge level but also show that the battery is being recharged. As the battery charges, the LED flashes and moves to the top of the charge indicator. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 36: Users Manual

    753/754 Users Manual gks9f.eps Figure 8. Battery Removal and Charger Use Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 37: Preserve Battery Life

    Auto Battery Save operates the same when the To avoid damage to the Product, use only the battery charger is used. Battery that comes with the Product, Fluke model BP7240, part number 4022220. To turn on the Auto Battery Save feature: Where ac power is available, the battery charger can be used to conserve battery power and energize the Product.
  • Page 38: Display Languages

    753/754 Users Manual Display Languages Display Intensity The Product shows information in five languages: There are two ways to increase display intensity: • • English Push . There are three levels of intensity when using this button. • European French •...
  • Page 39  or the Choices softkey to or s to save the settings and exit Setup mode. choose a setting for that parameter. For example, the display in Figure 10 is shown after you select Date Format. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 40: The Backlight

    753/754 Users Manual When the backlight dims, the Product also beeps. The Backlight Push  to change the backlight intensity from dim to bright Personalize the Product and back again.  shows at the top of the display when the Alphanumeric identifiers can be put into the Product to be backlight is active.
  • Page 41: Measure Mode

    If the range is locked, overrange inputs show on the display as - - - - - -. In Auto Range, out of range is shown as ! ! ! ! ! !. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 42: Electrical Parameter Measurement

    753/754 Users Manual Electrical Parameter Measurement When the Product is turned on, it is in the dc voltage measurement function. Figure 12 shows electrical measurement connections. To select an electrical measurement function from SOURCE or MEASURE/SOURCE mode, first push  for MEASURE mode: Push ...
  • Page 43 Documenting Process Calibrator Measure Mode – gks10f.eps Figure 12. Electrical Measurement Connections Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 44: Continuity Test

    Many ranges and types of pressure modules are available Short is shown on the display when the resistance between from Fluke. See “Accessories”. Before you use a pressure module, read its instruction sheet. The modules are different the Ω MEASURE jack and its common jack is less than in how they are used, zeroed, what types of process 25 Ω.
  • Page 45 • Use the pressure module only with gks11f.eps specified materials. See the printing on the Figure 13. Gage and Differential Pressure Modules pressure module or the pressure module instruction sheet for the acceptable material compatibility. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 46 753/754 Users Manual Connect a pressure module to the Product as shown in If necessary, the pressure display units can be changed Figure 14. The threads on the pressure modules accept to psi, mHg, inHg, inH2O, ftH2O, mH2O, bar, Pa, standard ¼...
  • Page 47 Documenting Process Calibrator Measure Mode Isolation Valve Gage Module Differential Module Tank gks12c.eps Figure 14. Pressure Measurement Connections Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 48: Temperature Measurement

    753/754 Users Manual If necessary, change between °C, °F, °R, and °K Temperature Measurement Temperature Units as follows: Thermocouple Use Push s. The Product supports thirteen standard thermocouples, Push the Next Page softkey twice. each identified with an alpha character: E, N, J, K, T, B, R, S, C, L, U, XK, or BP.
  • Page 49: Thermocouple Types Accepted

    -200 to 800 [1] American National Standards Institute (ANSI) device negative lead (L) is always red. [2] International Electrotechnical Commission (IEC) device negative lead (L) is always white. [3] Not an ANSI designation but a Hoskins Engineering Company designation. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 50 753/754 Users Manual Process Temperature Warning 30 V maximum to TC Miniplug gks12f.eps Figure 15. Temperature Measurement with a Thermocouple Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 51: Resistance-Temperature Detectors (Rtds)

    1000 Ω Platinum 0.00385 Ω/Ω/°C -200 to 630 9.035 Ω 0.00427 Ω/Ω/°C Cu10 (427) Copper -100 to 260 100 Ω 0.003916 Ω/Ω/°C Pt100 (3916) Platinum -200 to 630 [1] Per IEC 751-Standard [2] 10 Ω @ 25 °C Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 52 753/754 Users Manual To measure temperature where an RTD input is used: Caution To prevent possible damage to the Product, do If necessary, push  for MEASURE mode. not force a dual banana plug between any two Push t. jacks in the horizontal orientation. Doing so will damage the jacks.
  • Page 53 Push the Done softkey or s to exit Setup mode. If necessary, change between ITS-90 or IPTS-68 Temperature Scale in Setup mode. The procedure is the same as steps 1-7 above. gks14f.eps Figure 16. Correct Jumper Use Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 54 753/754 Users Manual Wire Wire Wire gks15f.eps Figure 17. Temperature Measurement with an RTD Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 55: Measurement Scale

    Push the Done softkey. pushed and you select a different scale mode. Percent of scale stays in effect until you change to a different measurement function or until you push the Scale softkey and select a different scale mode. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 56: Measure Or Source With Custom Units

    The measurement or source display can be setup to show The Fluke 700-IV Current Shunt is designed specifically for custom units. To do this, select a function, for example mV use with the 700 Series Documenting Process Products.
  • Page 57: Damping Measurements

     until SOURCE is shown. The Product damping method is a running average of the last eight must be in SOURCE mode to change any of the SOURCE measurements. Fluke recommends leaving damping on. parameters. Turning damping off may be useful when measurement...
  • Page 58 753/754 Users Manual Note To set the output value in the present source function, push  then enter the desired value and push . If sourcing frequency, respond when the Product asks you to select a zero-symmetric sine or To turn off sourcing completely, push  twice.
  • Page 59 Documenting Process Calibrator Source Mode Common Common gks16f.eps Figure 18. Electrical Source Connections Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 60: To 20 Ma Transmitter Simulation

    753/754 Users Manual 4 to 20 mA Transmitter Simulation The Product can be configured as a load on a current loop through the SOURCE mA function. In SOURCE mode, when  is pressed, the display prompts to select Source mA or Simulate Transmitter. When you Source mA the...
  • Page 61 Documenting Process Calibrator Source Mode DOCUMENTING PROCESS CALIBRATOR Loop Power Supply – gks17c.eps Figure 19. Connections to Simulate a 4 to 20 mA Transmitter Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 62: Supply Loop Power

    753/754 Users Manual Connect the Product in series with the instrument current Supply Loop Power loop as Figure 20 shows. To supply loop power: The Product supplies loop power at 26 V dc through an internal series resistance of 250 Ω. The setting supplies Push s for Setup mode.
  • Page 63 Documenting Process Calibrator Source Mode DOCUMENTING PROCESS CALIBRATOR TEST DC PWR – + – – gks18c.eps Figure 20. Connections to Supply Loop Power Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 64: Source Pressure

    Many ranges and types of pressure modules are available module. from Fluke, see “Accessories”. Before you use a pressure • Always apply correct torque between the module, read its instruction sheet. The modules are different...
  • Page 65 The pressure module MUST be Push  with the cursor on Pressure Units. zeroed before doing a task that sources or measures Select the pressure units with  or . pressure. Push . Push the Done softkey. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 66 753/754 Users Manual Pressure Module Hand Pump gks19c.eps Figure 21. Connections to Source Pressure Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 67: Thermocouple Simulation

    TC input/output. See Figure Temp.. After you record the external reference temperature, the Product corrects all voltages to adjust for this new reference junction temperature. If necessary, push  for SOURCE mode. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 68: Rtd Simulation

    753/754 Users Manual RTD Simulation Note See Table 8 for data about RTD (Resistance- Temperature Detector) types compatible with the Product. Connect the Product to the instrument under test as shown in Figure 23. The figure shows connections for two, three, or four-wire transmitters.
  • Page 69 Documenting Process Calibrator Source Mode DOCUMENTING PROCESS CALIBRATOR TEST DC PWR Color depends – on type of TC TC Miniplug – gks20c.eps Figure 22. Connections to Simulate a Thermocouple Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 70 753/754 Users Manual gks21f.eps Figure 23. Connections to Simulate an RTD Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 71: Source Temperature With A Hart Scientific Drywell

    If the Product will not set the The Product will begin to search for a drywell. If the drywell to temperatures within the drywell spec, refer to the Product shows Attempting connection for more than drywell manual to check the "High Limit" setting. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 72 753/754 Users Manual Null Modem Drywell Fluke Interface Cable Drywell (DB9) Fluke 3.5 mm Interface Cable Fluke Drywell (3.5 mm) gks99f.eps Figure 24. Source Temperature with Drywell Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 73: Source Scale

    Use the numeric keypad to record the 0% of scale value (100% Value). value (0% Value). Push . Push . Push the Done softkey. Use the numeric keypad to record the 100% of scale value (100% Value). Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 74: Step And Ramp The Output Value

    753/754 Users Manual Square root percent of scale is in effect until the Product To step the source value in % of scale, set the % of is changed to a different source function or the Scale scale value as given before in “Measurement Scale”.
  • Page 75: Ramp The Output

    1 V change in dc voltage or a change To Push the Stop Step softkey to stop the automatic in continuity status (Open or Short) from one ¼ second step function. interval to the subsequent interval. Push the Done softkey to continue normal operation. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 76 753/754 Users Manual To ramp (for example, sweep the source): See the applicable section earlier in this manual (for example, “Source Electrical Parameters”) and connect the Product to the test circuit. To automatically stop the Ramp function if a trip...
  • Page 77 Documenting Process Calibrator Source Mode DOCUMENTING PROCESS CALIBRATOR Trip Relay Voltage Trip Detect Output – Input 4 – 20 mA – Source mA gks22c.eps Figure 26. Check a Relay Output Trip Alarm Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 78: Simultaneous Measure/Source

    753/754 Users Manual Table 9 shows the functions that can be used at the same Simultaneous Measure/Source time when Loop Power is disabled. Table 10 shows the Use the MEASURE/SOURCE mode to calibrate or emulate functions that can be used at the same time when Loop a process instrument.
  • Page 79: Simultaneous Measure/Source Functions With Loop Power Disabled

    • Continuity • • • • • • • • • • • • • • • • • • 3W RTD • • • • • • 4W RTD • • • • • • Pressure Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 80: Simultaneous Measure/Source Functions With Loop Power Enabled

    753/754 Users Manual Table 10. Simultaneous MEASURE/SOURCE Functions with Loop Power Enabled Source Function Measure Function Ω dc V Freq Pressure • • • • • • dc V • • • • • • ac V • • •...
  • Page 81: Process Instrument Calibration

    Push  and  to select the thermocouple type. Push  to select then  to select Linear T source mode. Record a source value, for example 100 degrees, and then push . Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 82 753/754 Users Manual Push  for MEASURE/SOURCE mode. The display The display changes to the screen shown in Figure 29. changes to the screen shown in Figure 28. gks44s.bmp Figure 29. Process Instrument Calibration Screen 2 gks42s.bmp Figure 28. Process Instrument Calibration Screen 11.
  • Page 83 Documenting Process Calibrator Process Instrument Calibration DOCUMENTING PROCESS CALIBRATOR TEST DC PWR – – Original Circuit Wiring Power – Supply gks23c.eps Figure 30. Calibrate a Thermocouple Temperature Transmitter Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 84 753/754 Users Manual 12. More delay time can be input for the process instrument 15. The Test Strategy is the number of test points and to become stable than the Product’s usual settling time which test points are performed rising and falling in (about 2 seconds).
  • Page 85 If necessary, push Abort to exit from the calibration procedure. The tests start at the first test point, sources the correct temperature and measures the corresponding current from the transmitter. gks47s.bmp Figure 33. Error Summary Screen Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 86: Transmitter Adjustment

    753/754 Users Manual In the results summary test, failures are highlighted. An Push the Adjust softkey. The Product sources 0 % of adjustment is necessary in this example because three span (100 °C in this example) and shows these tests show failures. The failures were outside the softkeys: ±0.5 % tolerance that was selected.
  • Page 87: As Left" Test Run

    As Found calibration template is complete. • Measure/Source displays are in engineering units. • The measurement percentage is automatically corrected for the transmitter’s square-root response, and is used to calculate instrument errors. gks48s.bmp Figure 34. As Left Data Screen Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 88: Switch Calibration

    753/754 Users Manual Select the √ instrument procedure in a menu after you push the As Found softkey. Switch Calibration Switch B The procedure to calibrate a switch also uses the As Found Open Switch B and As Left calibration templates. Select the 1 Pt. Switch or Setpoint 2 Pt.
  • Page 89 14. Set the Tolerance to 0.5 psi. Push p for the pressure source function. 15. The next parameters, Deadband Min and Deadband Max, are optional. Do not set them in this example. These parameters would describe the minimum allowable size of the deadband. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 90 753/754 Users Manual 16. Push  to move through the choices to Set Trip 26. Push the Done softkey. Function to Trip Cont. 27. Push the As Left softkey to start the test again with the 17. Push the Done softkey.
  • Page 91: Transmitter Mode

    Transmitter mode is for diagnostic purposes Push the applicable function key for the process input. only. Use a completely charged battery. Do not use the Product in place of a transmitter for Push  for SOURCE mode. extended periods. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 92: Memory Operations

    753/754 Users Manual Push the applicable function key for the control output 15. To change the Transmitter mode parameters, push (for example,  or ). If the transmitter is connected Change Setup, and do the procedure in step 13 again.
  • Page 93 (Tag), instrument serial number (S/N), and Figure 36. operator name (ID), as shown in Figure 37. gks49s.bmp gks50s.bmp Figure 36. Saved Data Screen Figure 37. Additional Data Input Screen Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 94 753/754 Users Manual Record alphanumeric characters into the highlighted field The display shows an alphanumeric entry window. See with the optional bar code reader or the Product buttons. Figure 38. To record alphanumeric characters using the Product buttons, push  with the cursor on the necessary field to change (for example, Tag, above).
  • Page 95: Review The Memory

    See Figure 40. gks52s.bmp Figure 39. Memory Review Screen gks53s.bmp Figure 40. Data Log Parameters Screen Push  or  and  or the Go to Result softkey to view a result that was kept. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 96: Duration Limits

    753/754 Users Manual To log data: Table 11 gives an estimate of the limits for duration, assuming that no memory is being used for other purposes. If necessary, push  for MEASURE mode. Push the More Choices softkey twice. Table 11. Duration Limits Push the Log softkey.
  • Page 97 A long Push the Done softkey. The display changes to the logging duration can exceed the life of a screen shown in Figure 41. battery charge. gks54s.bmp Figure 41. Start Logging Screen Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 98: Record Min And Max Measurements

    753/754 Users Manual 10. Note the LOG annunciator next to MEASURE. Push Record Min and Max Measurements the Start Logging softkey to record data. You can set the display to record and show the maximum (max) and minimum (min) readings. Min and Max readings 11.
  • Page 99: Run A Preloaded Task

     to erase the memory: Push the Done softkey to start normal Product operation. • Results that have been kept • Min Max data • Log data sets A confirmation message is shown so that the memory is not accidently erased. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 100: Save To And Recall From The Registers

    753/754 Users Manual Save to and Recall from the Registers Use the Calculator to Set the Source Value When the Product is in calculator mode, the top half of the When you store to SOURCE, the Product shows a selection...
  • Page 101 Documenting Process Calibrator Quick Guide to Applications Measure V DOCUMENTING PROCESS CALIBRATOR DOCUMENTING PROCESS CALIBRATOR Circuit Ω 0 to 1V dc Input gks25c.eps gks26c.eps Figure 44.Voltage Drop Measurement Figure 43.Chart Recorder Calibration Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 102 753/754 Users Manual Measure DOCUMENTING PROCESS CALIBRATOR gks27c.eps Figure 45. Monitor AC Line Voltage and Frequency Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 103 Documenting Process Calibrator Quick Guide to Applications Measure Pressure Source mA DOCUMENTING PROCESS CALIBRATOR Loop Power Disabled Pressure Module gks28c.eps Figure 46. Current-to-Pressure (I/P) Transmitter Calibration Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 104 753/754 Users Manual DOCUMENTING PROCESS CALIBRATOR Measure mA Loop Power Disabled TEST DC PWR – – Original Circuit Wiring Power – Supply gks29c.eps Figure 47. Output Current of a Transmitter Measurement Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 105 Documenting Process Calibrator Quick Guide to Applications Measure Resistance DOCUMENTING PROCESS CALIBRATOR Source DOCUMENTING PROCESS CALIBRATOR Resistance Circuit gks30c.eps gks31c.eps Figure 48. Precision Resistor Measurement Figure 49. Resistance Source Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 106 753/754 Users Manual Measure Measure Continuity Frequency DOCUMENTING PROCESS CALIBRATOR DOCUMENTING PROCESS CALIBRATOR gks32c.eps gks33c.eps Figure 50. Checking a Switch Figure 51. Tachometer Examination Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 107 Documenting Process Calibrator Quick Guide to Applications Measure mA Source Pressure Loop Power Enabled Pressure Module Hand Pump gks34c.eps Figure 52. Analog and HART Pressure Transmitter Connection Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 108 753/754 Users Manual Measure mA DOCUMENTING PROCESS CALIBRATOR Output – Input 0 – 1mV, 0 – 10mV, 0 – 100mV, 1 – 5V, 0 – 1V, 0 – 10V Output 4 – 20 mA Input – Source mV or V gks35c.eps...
  • Page 109 Documenting Process Calibrator Quick Guide to Applications Measure Frequency DOCUMENTING PROCESS CALIBRATOR gks36c.eps Figure 54. Vortex Shedding Flowmeter Check Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 110 753/754 Users Manual – 2-Wire RTD= 3-Wire RTD= 4-Wire RTD= gks60.eps Figure 55. HART and Analog RTD Transmitter Connections Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 111 Documenting Process Calibrator Quick Guide to Applications – gks61.eps Figure 56. Analog and HART Thermocouple Transmitter Connections Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 112 753/754 Users Manual gks43.eps Figure 57. Transmitter HART- Comm Only Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 113: Communication With A Pc

    PC. A PC, Microsoft parts is available in the 75X Series Calibration  Windows, USB cable (supplied), and Fluke DPCTrack2 Manual available from the Fluke website. application software, or a qualified Fluke partner’s software are required. See the DPCTrack2 Users Manual Battery Replacement for further instructions.
  • Page 114: Calibration Data

    Be sure to pack the Warning Product securely, using the original shipping container if it is available. See “How to Contact Fluke” and the Warranty To avoid possible electric shock or personal Statement. injury, do not use the Product if it operates abnormally.
  • Page 115: Replacement Parts

    Documenting Process Calibrator User-Replaceable Parts Table 12. Replacement Parts Item Fluke Part Number Adjustable Quick-Release Strap 3889532 Input/Output Jack Decal 3405856 Tilt Stand 3404790 BP7240 Battery 4022220 USB Cable 1671807 BC7240 Power Supply/Battery Charger 4022655 Lens Cover 3609579 Alligator Clip Set-Extended Tooth...
  • Page 116: Accessories

    753/754 Users Manual • C700 Hard Carrying Case Accessories The Fluke accessories listed below are compatible with • BP7240 Li-Ion Battery the Product. For more information about these • TL series test leads accessories and their prices, contact a Fluke representative.
  • Page 117 Documenting Process Calibrator Accessories • FLUKE-700P27 300 psi / 0.01 • FLUKE-700P07 500 psi/0.01 • FLUKE-700P08 1000 psi/0.1 • FLUKE-700P09 1500 psi/0.1 • FLUKE-700PA3 5 psi/0.0001 • FLUKE-700PA4 15 psi/0.001 • FLUKE-700PA5 30 psi/0.001 • FLUKE-700PA6 100 psi/0.01 • FLUKE-700PV3 -5 psi/0.0001 •...
  • Page 118: Specifications

    753/754 Users Manual Specifications General Specifications All specifications apply from +18 °C to +28 °C unless stated otherwise. All specifications assume a 5-minute warmup period. Measurement specifications are valid only when Damping is turned on. When damping is turned off, or when the  annunciator is shown, floor specifications are multiplied by 3.
  • Page 119: Detailed Specifications

    Maximum Input Voltage: 300 V, IEC 61010 300 V CAT II Temperature coefficient: (0.001 % of reading + 0.001% of range) / °C (<18 °C or >28 °C) Normal mode rejection: >100 dB at 50 or 60 Hz nominal Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 120: Dc Voltage Measurement

    753/754 Users Manual DC Voltage Measurement % of Reading + Floor Range Resolution 1 Year 2 Year 0.00001 V 0.02 % + 0.00005 V 0.03 % + 0.00005 V ±3.00000 V ±30.0000 V 0.0001 V 0.02 % + 0.0005 V 0.03 % + 0.0005 V...
  • Page 121: Dc Current Measurement

    Open-circuit voltage: 5 V nominal Temperature coefficient: 3 % of specified accuracy / °C (<18 °C or >28 °C) Continuity Testing Tone Resistance Continuous tone <25 Ω May or may not get tone 25 to 400 Ω No tone >400 Ω Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 122: Frequency Measurement

    753/754 Users Manual Frequency Measurement Ranges Resolution 2 Year 1.00 Hz to 110.00 Hz 0.01 Hz 0.05 Hz 110.1 Hz to 1100.0 Hz 0.1 Hz 0.5 Hz 1.101 kHz to 11.000 kHz 0.001 kHz 0.005 kHz 11.01 kHz to 50.00 kHz 0.01 kHz...
  • Page 123: Dc Current Source

    Temperature Coefficient: (0.01 % of output +0.02 % of range / °C (<18 °C or >28 °C). When connected to mains, accuracy is not specified with conducted RF >1V, 8 to 15 MHz. Add 0.01 Ω when the excitation current is <1 mA. Add 0.0015 kΩ when the excitation current is <0.1 mA. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 124: Frequency Sourcing

    753/754 Users Manual Frequency Sourcing Specification Range 2 Year Sine Wave: 0.1 Hz to 10.99 Hz 0.01 Hz Square Wave: 0.01 Hz to 10.99 Hz 0.01 Hz Sine and Square Wave: 11.00 Hz to 109.99 Hz 0.1 Hz Sine and Square Wave: 110.0 Hz to 1099.9 Hz 0.1 Hz...
  • Page 125: Temperature, Thermocouples

    -210 to -100 -100 to 800 800 to 1200 -200 to -100 -100 to 400 400 to 1200 1200 to 1372 -250 to -200 -200 to 0 0 to 400 600 to 800 800 to 1000 1000 to 1820 Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 126 753/754 Users Manual Measure °C Source °C Range °C Type 1 Year 2 Year 1 Year 2 Year -20 to 0 0 to 100 100 to 1767 -20 to 0 0 to 200 200 to 1400 1400 to 1767 0 to 800...
  • Page 127 Instrument Operating Temperature: 0 to 50 °C for C and BP type thermocouples / -10 to 50 °C for all other types Normal Mode Rejection: 65 dB at 50 Hz or 60 Hz nominal For sourcing thermocouple voltages, accuracy is not specified in RF fields >1 V/m, 80 MHz to 700 MHz. Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 128: Temperature, Resistance Temperature Detectors

    753/754 Users Manual Temperature, Resistance Temperature Detectors Temperature, RTDs Degrees or % of Reading Measure °C Source °C Allowable Range °C Type (α) Excitation Source 1 Year 2 Year 1 Year 2 Year Current Current -200 to 100 0.07 °C 0.14 °C...
  • Page 129: Loop Power

    Pt(3916): JIS C 1604, 1981 Pt(3926), Cu(427), Ni(672): Minco Application Aid #18 Loop Power Open Circuit Loaded Circuit 18 V minimum at 22 mA 26 V ±10 % Short circuit protected to 25 mA Output Resistance: 250 Ω nominal Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...
  • Page 130 753/754 Users Manual Fluke-Direct info@Fluke-Direct.ca 1.800.868.7495...