Fluke 753 User Manual page 89

Documenting process calibrator
Hide thumbs Also See for 753:
Table of Contents

Advertisement

To do a pressure switch test: The switch in this example
sets at a high limit of 10 psi. The set condition is a closed
switch contact. For pressure switches, use the Manual Test
selection. For switches where sourcing pressure is not
necessary, use the Auto Test selection to do the test.
1.
Connect the test leads between the pressure switch
contact output and the mA Ω RTD (center) jacks on the
Product.
2.
Connect the pressure module to the Product, and
connect a pressure line to the switch. Keep the
pressure line vented to atmosphere.
3.
If necessary, push  for MEASURE mode.
4.
Push q for the continuity measure function.
5.
Push  for SOURCE mode.
6.
Push p for the pressure source function.
info@Fluke-Direct.ca
7.
Push  to zero the pressure module.
8.
Push .
9.
Push the As Found softkey.
10. Highlight 1 Pt. Switch Test from the menu and push
.
11. Push  to modify the parameters for Setpoint 1.
12. Make these selections:
Setpoint 1 = 10.000 psi
Setpoint Type = High
Set State = Short
13. Push the Done softkey.
14. Set the Tolerance to 0.5 psi.
15. The next parameters, Deadband Min and Deadband
Max, are optional. Do not set them in this example.
These parameters would describe the minimum
allowable size of the deadband.
Fluke-Direct
.ca
Documenting Process Calibrator
Process Instrument Calibration
1.800.868.7495