Detailed Specifications; Dc Mv Measurement - Fluke 753 User Manual

Documenting process calibrator
Hide thumbs Also See for 753:
Table of Contents

Advertisement

Double Insulation Creepage and Clearance ....................... Per IEC 61010-1
Installation Category ............................................................. 300 V CAT II
Design Standards and Compliance ..................................... EN/IEC 61010-1:2010, CAN/CSA C22.2 No. 61010-1-04, ANSI/UL 61010-
EMI, RFI, EMC ........................................................................ EN 61326-1:2006
RF Fields ................................................................................ Accuracy for all functions is not specified in RF fields >3 V/m

Detailed Specifications

Specifications valid after a 5-minute warmup.
Specifications are valid to 110 % of Range with the following exceptions: 300 V dc measure, 300 V ac measure, 50 kHz measure and
source, 22 mA source and simulate, 15 V dc source, and temperature measure and source which are valid to 100 % of range.

DC mV Measurement

Range
0.001 mV
±100.000 mV
Input Impedance: >5 MΩ
Maximum Input Voltage: 300 V, IEC 61010 300 V CAT II
Temperature coefficient: (0.001 % of reading + 0.001% of range) / °C (<18 °C or >28 °C)
Normal mode rejection: >100 dB at 50 or 60 Hz nominal
info@Fluke-Direct.ca
1:2004
Resolution
0.02 % + 0.005 mV
Fluke-Direct
Documenting Process Calibrator
% of Reading + Floor
1 Year
.ca
1.800.868.7495
Detailed Specifications
2 Year
0.03 % + 0.005 mV
107