Measurement Scale
This feature scales measurements in accordance with a
applicable process instrument's response. Percent of scale
works for linear-output transmitters or square-law
transmitters such as differential pressure transmitters that
report flow rate.
Linear-Output Transmitters
1.
If necessary, push for MEASURE mode.
2.
Select a measurement function (, , h, q, t,
or p) as previously described.
3.
Push the Scale softkey.
4.
Select % from the list.
5.
Use the numeric keypad to record the 0% of scale
value (0% Value).
6.
Push .
7.
Use the numeric keypad to record the 100% of scale
value (100% Value).
8.
Push .
9.
Push the Done softkey.
Percent of scale stays in effect until you change to a
different measurement function or until you push the Scale
softkey and select a different scale mode.
info@Fluke-Direct.ca
Square-Law Process Variables
When you select √ within scaling, the Product takes the
square root of its input and shows the measurement in
percent. For example, when the Product is connected to the
output of a delta-pressure transmitter, the Product indication
is in proportion to flow rate.
1.
If necessary, push for MEASURE mode.
2.
Select a measurement function (, , h, q, t,
or p) as previously described.
3.
Push the Scale softkey.
Select √ from the list.
4.
5.
Use the numeric keypad to record the 0% of scale
value (0% Value).
6.
Push .
7.
Use the numeric keypad to record the 100% of scale
value (100% Value).
8.
Push the Done softkey.
Square root percent of scale stays in effect until you change
to a different measurement function or the Scale softkey is
pushed and you select a different scale mode.