Measurement Scale
This feature lets you scale the measurements in
accordance with a particular process instrument's
response. Percent of scale works for linear-output
transmitters or square-law transmitters such as
differential pressure transmitters that report flow rate.
Linear-Output Transmitters
1. If necessary, press
2. Select a measurement function (
,
,
, or
3. Press the
softkey.
Scale
4. Select
from the list.
% scale
5. Use the numeric keypad to enter the 0% of scale
value (
).
0% Value
6. Press
.
7. Use the numeric keypad to enter the 100% of
scale value (
100% Value
8. Press
.
for MEASURE mode.
,
,
) as previously described.
).
Documenting Process Calibrator
9. Press the
softkey.
Done
Percent of scale remains in effect until you change to
another measurement function or press the
softkey and select another scale mode.
Using Measure Mode
Scale