Upper jaw
Lower jaw
Upper jaw
Lower jaw
Exposure time in
seconds with:
60 kV
70 kV
Freely programmed
values
62 14 717 D3507
D3507.201.01.13.02 08.2015
PLUS
4.8.3
Pre-programmed exposure times for films of sensitivity
class E and with a 300 mm (12") FHA cone
Exposure times film E cone 300 WW
0,12
0,16
0,20
0,06
0,08
0,10
NOTICE
For film of sensitivity class F: Set the exposure time one level lower with
the minus button.
For films of sensitivity class D: Set the exposure time four levels higher
with the plus button.
Using a film holder: Set the exposure time one or two levels higher with
the plus button.
4 Controls and functional elements
0,25
0,32
0,40
0,12
0,16
0,20
4.8 Exposure times
0,50
0,64
0,25
0,32
25