3 Technical description
3.1 Technical data
14
Source-skin distance:
Class I device
Degree of protection against
electric shock:
Degree of protection against
ingress of water:
Year of manufacture:
Operating mode:
X-ray tubes:
Nominal continuous power
rating of the X-ray tube:
Power rating of X-ray tube
(70kW/7mA):
Anode material:
Anode angle:
Exposure parameters for
determining leakage
radiation:
Leakage radiation at 1 m
distance:
Transport and operating conditions:
Transport and operating conditions
Transport and storage
temperature:
Air humidity:
Operating conditions as
specified in IEC 60601-1:
Operating altitude:
Sirona Dental Systems GmbH
Operating Instructions HELIODENT
FHA 200 mm (8") - standard or 300 mm
(12")
Type B device
Ordinary equipment (without protection
against ingress of water)
(on the rating plate)
Continuous operation
Petrick P470/8.35/12G
26 W
490 W
Tungsten
12°
0.12 mA / 70 kV
< 0.25 mGy/h
-40°C – +70°C (-40°F – 158°F)
10% – 95%
Ambient temperature +10 °C – +40 °C
(50 °F – 104 °F)
Relative humidity: 30% – 75%
≤ 3000 m
D3507.201.01.13.02 08.2015