Sirona Dental Systems GmbH
Operating Instructions SINIUS / SINIUS CS
63 22 726 D3561
D3561.201.01.03.02 05.2012
5.3.4
Care for, disinfect and sterilize the treatment instruments
Care for, disinfect and sterilize the treatment instruments
Treatment instruments
5.3.4.1
Treatment instruments
The procedures required for the following treatment instruments are
described in the following separate operating instructions:
List of treatment instruments, without SL motor and implant and HF surgery
● BL motor
● BL ISO S/C motor
● Straight and contra-angle handpieces in various versions
● High-speed handpieces
● SPRAYVIT M multifunctional syringe
● SIROSONIC TL scaler
5.3.4.2
Cleaning and disinfecting/sterilizing the components of the
ApexLocator
Cleaning and disinfecting/sterilizing the components of the ApexLocator
Care and cleaning agents
NOTICE
Approved care and cleaning agents
Use only care and cleaning agents which have been approved by Sirona
(see "Care and cleaning agents" [ → 170]).
Disconnect the mucosal electrode from the connection cable. The metal
hook can be sterilized, and the connection cable can be disinfected by
wiping.
134°C 273,2°F
3 min
The file clamp for manual measurement can be sterilized with the
connection cable.
134°C 273,2°F
3 min
The apex adapter and its connection cable can be disinfected by wiping.
ApexLocator, no disinfectant between contacts
IMPORTANT
To guarantee conductivity, no disinfectant must be allowed to penetrate
the electrical contacts.
5 Care, cleaning and maintenance by the practice team
5.3 Instruments and instrument hoses
187