Table 3.2 Troubleshooting Chart
Problem
Rotor cannot achieve
set speed
Door will not open
Displays are blank
Chamber does not
reach selected
temperature
Chamber icing

Retrieving Your Sample in Case of Power Failure

If facility power fails you will have to restart the run when the power is restored. In the event of an
extended power failure, it may be necessary to trip the door-locking mechanism manually to
remove the rotor and retrieve your sample.
1
Turn the power off and disconnect the power cord from the main power source.
• No displays or LEDs are lit.
PN B07493AD
Problem/Result
Line voltage below rating
Electrical failure
Motor failure
Rotor spinning
Power not on
Source power failure
Power not on
Electrical failure
Centrifuge cannot
maintain selected
temperature for rotor in
use at speed selected
Excess humidity in
chamber
Condensation build-up
between runs
Retrieving Your Sample in Case of Power Failure
Recommended Action
Have a qualified service person measure line
voltage while the instrument is operating.
Make sure the power cord is securely connected;
call Beckman Coulter Field Service.
Call Beckman Coulter Field Service.
Wait until the rotor stops.
Plug in the power cord; turn power on.
See
Retrieving Your Sample in Case of Power
Failure, below.
Plug in the power cord; turn power on.
Make sure the power cord is securely connected;
call Beckman Coulter Field Service.
• Precool or prewarm rotors before running at low
or high temperatures.
• Precool rotor chamber by running a 30-minute
cycle at the required temperature with the
speed set at 2000 RPM.
• Make sure the air intake vent is clear.
• If using multiwell-plates for runs exceeding
2 hours in a warm and/or humid environment,
speed reduction may be required to maintain
low sample temperature.
Wipe out moisture from the chamber and the
chamber gasket before each run.
• Leave door open between runs.
• Set temperature to a setting higher than
ambient temperature.
• Turn centrifuge power off.
3
Troubleshooting
3-5