Power Module R-net

Removal
1. Raise the seat to the highest position. If the chassis is equipped with fixed seat post, see Seat height — Fixed seat
post, Page 96. If the seat lift does not work normally because the batteries are discharged or the actuator is defec-
tive, the seat can be raised/lowered manually. See Manual operation of Electric Seat lift, Page 41.
2. Switch off the main power switch on the control panel.
3. Put the circuit breaker in the "OFF" position. It is accessed through a hole in the chassis cover. See 78 Circuit
breaker/Battery isolator., Page 54.
4. Remove the seat lift cover and the chassis cover, see page Covers, Page 14.
5. Lift the R-net controller out of its holder, see fig. 68
6. Disconnect the electrical connections to the R-net controller, being attentive to their placement: see fig. 69.
Figure 68. The R-net controller sits in its container without fastening screws.
Assembly
Assemble in reverse order.
Note the location of the cable connections, see fig. 69 Power Module cable connections (Rnet PM90).
50
Figure 69. Power Module cable connections (Rnet PM90).