7.7

Error Codes

Common Alarms
ERROR
TRUE MEANING
CODE
256
Warning - Battery
less than 25%
512
Warning - Battery
less than 15%
768
Warning - No
breath found
1024
Warning - Battery
less than 10%
1280
Warning - Breath
rate high
1536
Alarm - Unit Temp
HI/LO
1792
Warning - Start
Temp HI/LO
2048
Alarm - Battery
Temp HI/LO
2049
Alarm - Battery
Temp HI/LO
2304
Warning - Low O2
Purity
2560
Alarm - Low O2
Purity
2816
Corrupt Scheduler IDF (interrupt driven framework) scheduler was not able to execute all scheduled events within 1 msec timer period.
Part No 1176122
POSSIBLE REASON/SOLUTION
Charge battery, see user manual
Charge battery, see user manual
Check cannula
Check for no leaks at outlet filter
Charge battery, see user manual
Reduce activity
Have patient breathe at a lower rate
Check ambient temp and blockage of air inlet/outlet
Check ambient temp and blockage of air inlet/outlet
Check ambient temp and blockage of air inlet/outlet
Check ambient temp and blockage of air inlet/outlet
Check ambient temp and blockage of air inlet/outlet
Check ambient temp and blockage of air inlet/outlet
If excess of 1 msec was allowed, the operation of the Solo2 could become unstable
7 TROUBLESHOOTING AND MAINTENANCE
27
Invacare® Solo
™ Portable Oxygen Concentrator
2