14 P.C. BOARD
14 P.C. Board
CAUTION
Before handling the P.C. Board, ensure you are properly grounded to prevent static discharge and damage to the components of the P.C.
Board. A static strap MUST be worn and properly grounded using an alligator clip. Electrical conduit or a water pipe can bue used if a
ground is unavailable. Care should be taken to ensure that the alligator clip contacts bare metal surface.

14.1 Removing the P.C. Board

For this procedure refer to Figures 1, 2, 3 and 4.
Required tools: Torx® T15 screwdriver
1.
Disassemble the Solo2. Refer to Accessing Internal Components on page 40.
2.
Disconnect the tubes from the P.C.Board side only.
3.
Disconnect the electrical connections from the P.C. Board side only.
4.
With the wires and tubes disconnected, remove the screws holding the P.C. Board to the compressor sound box.
Invacare® Solo2™ Portable Oxygen Concentrator
46
Part No 1176122