Invacare Solo2 TPO100B Internal Repair Manual page 9

Portable oxygen concentrator
Hide thumbs Also See for Solo2 TPO100B:
Table of Contents

Advertisement

IPX1
IPX2
NO AP/APG
POWER INPUT:
EXTERNAL POWER SUPPLY INPUT:
AC POWER SUPPLY:
DC POWER SUPPLY:
SOUND LEVEL:
ALTITUDE:
OXYGEN CONCENTRATION:*
*BASED ON AN ATMOSPHERIC PRESSURE OF
14.7 PSI
(101 KPA) AT 70°F (21°C) NOMINAL
CONSERVER TRIGGER SENSITIVITY:
FACTORY SET - NO ADJUSTMENT
PRESSURE ACTIVATED
Part No 1176122
Recycle
DO NOT use oil or grease
Keep dry in transport or storage
Protected against dripping water in upright position (AC and DC power input)
Protected against dripping water in upright and tilted positions (battery power only)
Not suitable for use in the presence of a flammable anaesthetic mixture
24 VDC @ 7.5 Amp or 11-16 VDC @ 10.0 Amp max (12.6 VDC Nom.)
120 VAC, 50/60 Hz @ 2.5 amps, 230 VAC, 50/60 Hz @ 1.25 amps
11-16 VDC, (12.6 VDC Nom. @ 10.0 Amp max)
< 40 dBA weighted @ 2 LPM continuous and all pulse settings (1-5)
Up to 10,000 ft (3046 m) above sea level
Titration recommended for use above 10,000 ft (3046 m)
87% to 95.6%, after initial warm-up period (approximately 5 minutes) at all flow rates
< 0.20 cmH
O max pressure drop
2
All settings - using 7 ft (2.1 m) cannula
9
Invacare® Solo2™ Portable Oxygen Concentrator
2 OVERVIEW

Advertisement

Table of Contents
loading

Table of Contents