11 ACCESSING INTERNAL COMPONENTS
11 Accessing Internal Components
WARNING
Ensure the battery pack and or the external power supplies are removed from the unit before attempting to disassemble the unit.
Required tools: Torx® T15 screwdriver, 1/8" hex bit driver

11.1 Disassembly

1.
Remove and retain the Indication Labels shown in Detail "A".
2.
Remove the 1/8" hex head screws holding the handle to the unit and
save for later reinstall.
3.
Remove and retain the #6-19 x .50 Plastite screws in Detail "B"
using the Torx™ T15 screwdriver from the control panel.
4.
Lay the unit face down as shown in Detail "C".
5.
Remove and retain the #6-19 x .50 Plastite screws in Detail "C"
using the Torx™ T15 screwdriver.
6.
While holding onto the front and back shrouds, flip the unit over
onto the back shroud as seen in Detail "D" ensuring the ribbon
cable is not pulled apart from the unit.
DETAIL "B"
Control Panel Screws
Invacare® Solo2™ Portable Oxygen Concentrator
DETAIL "C"
Control Panel
Top Shroud
40
DETAIL "A"
Indication Labels
DETAIL "D"
Ribbon Cable
Part No 1176122