Invacare Solo2 TPO100B Internal Repair Manual page 10

Portable oxygen concentrator
Hide thumbs Also See for Solo2 TPO100B:
Table of Contents

Advertisement

2 OVERVIEW
CONSERVER TRIGGER DELAY:
FACTORY SET - NO ADJUSTMENT
CONSERVER BOLUS DELIVERY:
FIXED MINUTE VOLUME
CONSERVER BREATH RATE CAPACITY: Up to, and including, 35 BPM without reduction of bolus minute volume
CONTINUOUS FLOW SETTINGS:
MAXIMUM RECOMMENDED
FLOWRATE WITH 7KPA (1.01 PSI)
BACKPRESSURE:
PRESSURE RELIEF ACTIVATION:
MAX OUTLET PRESSURE @ 3.0 LPM
CONTINUOUS:
DIMENSIONS: (WITHOUT CART)
WEIGHT: (UNIT WITHOUT CART)
BATTERY DURATION:
(TIMES ARE APPROXIMATE)
Invacare® Solo2™ Portable Oxygen Concentrator
35 mSec max
Nominal value - using 7 ft (2.1 m) cannula
Setting 1: 400 cc
Setting 2: 800 cc
Setting 3: 1200 cc
Setting 4: 1600 cc
Setting 5: 2000 cc
Total cc's ± 75 cc delivered per minute
0.5 through 3.0 LPM @ 0 psi, 0.5 LPM increments, All settings are ± 0.2 LPM
(2.0 LPM max w/Ext DC Power Cable)
3.0 LPM
20 psi ± 5 psi (137.8 kPa ± 34.5 kPa)
12.0 psig
16.5 in high x 11 in wide x 8 in deep (41.9 cm high x 27.9 cm wide x 20.3 cm deep)
< 20 lbs (9.09 kg)
Pulse Mode
Setting 1 = 4.5 hrs
Setting 2 = 3.5 hrs
Setting 3 = 3.0 hrs
Setting 4 = 2.5 hrs
Setting 5 = 2.5 hrs
10
Continuous Mode
Setting 0.5 LPM = 4.5 hrs
Setting 1 LPM = 3.5 hrs
Setting 2 LPM = 2.5 hrs
Setting 3 LPM = 1.5 hr
Part No 1176122